Emerson Rosemount 644 Reference Manual page 75

Temperature transmitter
Hide thumbs Also See for Rosemount 644:
Table of Contents

Advertisement

Reference Manual
00809-0200-4728
4.3.1
Sensor connections
The transmitter is compatible with a number of RTD and thermocouple sensor types.
Sensor wiring
transmitter. To ensure a proper sensor connection, anchor the sensor lead wires into the
appropriate captive terminals and tighten the screws.
Sensor wiring
Emerson provides a 4-wire sensors for all single element RTDs.
You can use these RTDs in 3-wire configurations by leaving the unneeded leads
disconnected and insulated with electrical tape.
Figure 4-2: HART Head Mount and Field Mount
Single input wiring
Dual input wiring
Figure 4-3: HART Rail Mount and F
1
2-wire
RTD and Ω
Emerson.com/Rosemount
shows the correct input connections to the sensor terminals on the
2 3 4
1
2
3-wire
RTD and Ω*
Fieldbus/PROFIBUS
OUNDATION
3
4
1
2
4-wire
RTD and Ω
Electrical Installation
March 2021
+
3
4
1
2
3
T/C
and mV
_
4
75