3.4.2
Mount the transmitter
Mount the transmitter at a high point in the conduit run to prevent moisture from draining
into the transmitter housing.
The Rosemount 644 Head Mount installs:
In a connection head or universal head mounted directly on a sensor assembly.
Apart from a sensor assembly using a universal head.
To a DIN rail using an optional mounting clip.
The Rosemount 644 Field Mount installs in a field mount housing, directly mounted on a
sensor or apart from a sensor assembly using an optional bracket.
The Rosemount 644 Rail Mount attaches directly to a wall or to a DIN rail.
Mounting a Rosemount 644 Head Mount to a DIN rail
To attach a head mount transmitter to a DIN rail, assemble the appropriate rail mounting
kit (part number 00644-5301-0010) to the transmitter as shown in
procedure under
Figure 3-5: Assembling Rail Clip Hardware to a Rosemount 644 Transmitter
A. Mounting hardware