Oxley Creek WWTP Water Reclamation
(Probe LU Manual) Vendor Manual
Write locking
Write locking prevents any changes to parameters, either via PDM or the handheld
programmer, but still allows access to the device.
Values
1.
Go to
1. dentification M enu
2.
Press
3.
Key in
4.
Press
mode.
Remote operation enable
Remote Operation can be enabled or disabled via the handheld programmer.
Values
To change the setting:
•
Open
•
Press
1. Identification
•
To enable programming, set REMLOCK to . To disable programming, enter
7ML19985JB02
1. 3. Configuration
1.3.5. Lock
to open Edit mode.
(unlock value) to enable parameter changes.
again, to accept the change and return to Navigation
Remote operation enabled.
Remote operation disabled.
Menu, then scroll down to CONFIG.
to open the Config Menu, then scroll down to REMLOCK.
1.2. Configuration
1.2.2. Remote Lock
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
Doc Id: VM11
Active Date: 15/12/2006
Enables parameter changes.
Disables parameter changes.
Page 45