Fluke 9009 User Manual page 29

Hide thumbs Also See for 9009:
Table of Contents

Advertisement

Problem
Temperature readout is not the actual
temperature of the well
Troubleshooting Problems, Possible Causes, and Solutions
Table 4. Troubleshooting (cont.)
Possible RF energy emission. With the
Product stable, slowly rotate the Product. If no
change occurs, the Product may need to be
calibrated. If so, contact a service center. If the
display changes more than twice the normal
display deviation, another Product in the area
could be emitting RF energy. Move the Product
to a different location and rotate the Product
again. If the temperature is correct in this new
area or deviates differently than the first are, RF
energy is present in the room. If you must
perform the test in the affected area, use the
comparison test to eliminate any possible errors.
Industrial Dual-Block Calibrator
Possible Causes and Solutions
25

Advertisement

Table of Contents
loading

Table of Contents