Power Line Connection Information - Siemens ARCADIS Varic Start-Up Instructions

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

22

Power line connection information

The delivery state of the system with respect to the power line connection values is indi-
cated by the line voltage label or in the operating instructions.
Any necessary adjustments to local conditions must be performed in accordance with the
service instructions/the circuit diagram of the system.
The system leakage current measurement guidelines must be observed.
WARNING
ARCADIS Varic
Electrical voltage!
Non-compliance can lead to severe bodily injury and even death.
The internal uninterruptible power supply of the system
(UPS) provides several components with line voltage -
even when the system is switched off or the system
power plug is disconnected.
Before the system is programmed to the local line volt-
age/line frequency, it must be disconnected from the
power supply (the power plug must be disconnected)
and the voltage output plug of the UPS must be discon-
nected as well.
SPR2-310.815.01.07.02
12.06
General information
Page 22 of 64
CS PS SP
1.6
Siemens AG
Medical Solutions

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents