Chapter 3 No-Picture Problem - Eizo RadiForce RX370 Instructions For Use Manual

Color monitor
Table of Contents

Advertisement

Chapter 3 No-Picture Problem

Problem
1. No picture
• Power switch indicator does not light up
• Power switch indicator lights up: Green
• Power switch indicator lights up: Orange
• Power switch indicator blinks: Orange,
Green
2. The message below appears.
• This message appears when no signal is
input.
Example:
• The message indicates that the input signal
is out of the specified frequency range.
Example:
Possible cause and remedy
• Check whether the power cord is connected properly.
• Turn the main power switch on.
• Touch
.
• Turn off the main power, and then turn it on again.
• Increase "Brightness", "Contrast", or "Gain" in the Setting
menu. For details, refer to the Installation Manual (on the
CD-ROM).
• Turn off the main power, and then turn it on again.
• Switch the input signal. For details, refer to the Installation
Manual (on the CD-ROM).
• Move the mouse or press any key on the keyboard.
• Check whether the PC is turned on.
• Check whether the signal cable is connected properly.
Connect to
when selecting "DisplayPort 1", and to
when selecting "DisplayPort 2" in the input signal.
for output when a daisy-chain connection is set up.
• Turn off the main power, and then turn it on again.
• Connect via the signal cable specified by EIZO. Turn off the
main power, and then turn it on again.
• If the signal cable is connected to DisplayPort 1 (
switching the DisplayPort version. For details, refer to the
Installation Manual (on the CD-ROM).
This message appears when the signal is not input correctly
even though the monitor is functioning properly.
• The message shown left may appear, because some PCs do
not output the signal immediately after power-on.
• Check whether the PC is turned on.
• Check whether the signal cable is connected properly.
Connect the signal cables to the connectors of the
corresponding input signal.
• Switch the input signal. For details, refer to the Installation
Manual (on the CD-ROM).
• If the signal cable is connected to DisplayPort 1 (
switching the DisplayPort version. For details, refer to the
Installation Manual (on the CD-ROM).
• Check whether the signal cable is properly connected.
Is the signal cable connected to
DisplayPort signal?
is used for output when a daisy-chain
connection is set up.
• Turn off the main power, and then turn it on again.
• Check whether the PC is configured to meet the resolution
and vertical scan frequency requirements of the monitor (see
"4-2. Compatible Resolutions" (page 23)).
• Reboot the PC.
• Select an appropriate setting using the graphics board's
utility. For more information, refer to the User's Manual of the
graphics board.
), try
), try
or
for inputting the
Chapter 3 No-Picture Problem
is used
21