Sterilizing Parameters

Item to be
sterilized
Rubber
Dressings
Utensils
Apparatus
Liquor
Dishware
Use calibration offset function
Calibration offset is a function which corrects the difference between the temperature in furnace and
that of controller (sensor temperature) if arises. The function parallel corrects the difference either to
the plus or minus side within the whole temperature range of unit. The function can be set or
cancelled by the
① Start operation with the target setting temperature. Check the temperature in furnace (temperature of
sample) with a thermograph after it is stabilized.
② Check the difference between the setting temperature and that in furnace (temperature of sample).
③ Press the
key. Select the character "SC", which indicates the calibration offset, using the "▲▼",
and then press the
④ Input the difference using the "▲▼" and then press the
The setting range of offset correction temperature is +20 ℃ to plus side and -20 ℃ to minus side
respectively.
Necessary heat preservation
time
15
30-45
15
10
20-40
20
key.
key.
Relative steam
temperature(℃)
121
121-126
121-126
121-126
121-126
121-126
Corrected temp. to plus side
Current temperature
Corrected temp. to minus side
key. This completes the setting.
14
Operation Method
Relative steam
pressure(MPa)
0.1-0.105
0.105-0.14
0.105-0.14
0.105-0.14
0.105-0.14
0.105-0.14