Installation Procedure

1
Set the attached drain board onto the bottom
surface inside the chamber
• The drain board stabilizes the sterile samples
inside the chamber as well as protects the
heater and sensor. Make sure to set it.
2
Close the drain valve
• Close the drain valve at the bottom on the
right side face of main unit. Water leak
occurs
if not fully closed, which may cause the burn
injury no-load (water) operation.
The drain valve and lead it to the draining
site.
3
Pour water into the chamber
• Before setting the sterile samples, pour water into the chamber to the water level gauge (notch)
position.
• Insufficient water may cause the no-load (water) operation. Check the water level every time
before operation. Refill it before the level becomes too low. Water is required to be poured at
dissolution operation, as well as sterilizing operation.
• When the water level lowers, the equipment cuts off the heater. Depending on the conditions of
equipment, however, the detection requires too much time, which may cause the heater
deterioration. Refill water before the water level becomes too low.
Refer to the right table for the quantity of water to be refilled.
4
Use distillated or purified water for sterilizing water
• Fill distillated or purified water inside the chamber. Tap water may be used. Calculus generate
inside the chamber when tap water is used. Frequent cleaning is therefore required.
• Do not use well water. It may cause the corrosion or dirt inside the chamber.
5
Set the sterile samples
• Set the samples to the chamber, putting them into the attached rack or cast (sold separately).
• Put the sample or sterilization bag into the chamber so they should not block or cover the
sensor inside the chamber, exhaust outlet and end connection to pressure gauge. If they are
blocked or covered, the vapor cannot be discharged and the equipment cannot be operated
correctly. Do not spill the samples when taking them out or putting them into the chamber. The
failure in piping system, bad smell or dirt may result in.
• In case liquid such as medicinal solution or medium is sterilized, the amount of liquid should be
60% or less of the capacity of container. They may be boiled over if too much quantity is
supplied.
• Widely open the opening of sterilization bag when used. If it is closed, the samples are
insufficiently sterilized.
Before Using This Unit
8
Set the drain board
horizontally.
The water level
indicator board