DX-D 300 | Introduction | 47
WARNING:
The DR Detector is not intended to be used as a primary barrier
to X-rays. The user is responsible for ensuring the safety of the
operator, bystanders, and the subjects being radiographed.
CAUTION:
Strictly observe all warnings, cautions, notes and safety
markings within this document and on the product.
CAUTION:
All Agfa medical products must be used by trained and qualified
personnel.
CAUTION:
Avoid unnecessary dose by checking before exposure if the DR
Detector Switch displays the name of the DR Detector that is
being used and if the status of the DR Detector is ready for
exposure.
WARNING:
There is a risk of collision or crushing for patients, operating
staff, unit and objects, caused by unit movements which could
be released by inadmissible actuation of operating elements by
patients.
WARNING:
System unavailability due to hardware or software failure. If the
product is used in critical clinical workflows, a backup system
has to be foreseen.
CAUTION:
Before moving the U-arm out of horizontal position, check that
no objects are lying on the bucky that can fall off.
CAUTION:
Damaged grid. Reduced image quality. Please handle the grids
with special care.
CAUTION:
When inserting the scattered radiation grids, it is essential that
the grid corresponds to the intended source-image-distance
(SID) to which the grid is focussed. Because of the focussing of
the grids, the tube unit must be centered onto the bucky.
CAUTION:
Excessive ambient temperature may impact performance of DR
Detectors and cause permanent damage to the equipment. Refer
to the related user manual for environmental conditions for the
DR detector. If ambient temperature and humidity is outside the
0172E EN 20181116 1653