Solimotion FTR16 Table of contents Table of contents About this document �������������������������������������������� 4 7.4.5 Reset to factory settings ......22 Function test ............. 22 Symbols ............... 4 1.1.1 Safety symbols ..........4 Commissioning ����������������������������������������������������22 1.1.2 Electrical symbols ........4 Function check ............22 1.1.3...
About this document Solimotion FTR16 About this document 1�1 Symbols 1�1�1 Safety symbols This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
‣ The operator is responsible for interference-free operation of the device. Conversions to the device Unauthorized modifications to the device are not permitted and can lead to unforeseeable dangers. ‣ If, despite this, modifications are required, consult with Endress+Hauser. Endress+Hauser...
It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser con- firms this by affixing the CE mark to the device.
• Enter the serial number on the nameplate into the Endress+Hauser Operations App or use the Endress+Hauser Operations App to scan the 2-D matrix code (QR Code) on the nameplate 4�2�1...
Mounting Solimotion FTR16 4�3�2 Transporting the device Transport the device to the measuring point in the original packaging. Mounting 5�1 Mounting conditions Minimization of application-specific influences ‣ Vibration effects → 34 5�1�1 Mounting position • Any mounting position • Small angle α can increase signal quality.
Solimotion FTR16 Mounting 5�1�3 Mounting dimensions 90 (3.54) 110 (4.33) 20 (0.79) 40 (1.57) 55 (2.17) 110 (4.33) 28 (1.10) (2.76) 0000000012 4 Mounting dimensions. Unit of measurement mm (in) 5�2 Mounting the device 5�2�1 Mounting in contact with the process The FTR16 is screwed directly into the process (for example existing threads or vessel sleeves) with its process connection (standard threads G 1 and G 1½...
Page 10
Mounting Solimotion FTR16 Mounting with connection thread 2 ³⁄₈" 0000000061 6 Mounting with connection thread 1½ NPT G 1 / G 1½ Screw in conical (A) or cylindrical (B) connection thread. Align the housing of the electronics. Fix the housing in place.
Solimotion FTR16 Mounting When using the G 1½ process connection (standard thread according to ISO 228-1, hexagon SW55) and using the optional counternut (→ 26), the device can be mounted flush particularly easily, as it is a cylindrical thread. Mounting with welding sleeve G 1 (A) ‣...
Page 12
Mounting Solimotion FTR16 Mountinginfrontofmicrowavepermeablesightglassfitting • Observe maximum temperature (→ 34) • Suitable sight glass fitting → 29 0000000023 9 Mounting in front of microwave permeable sight glass fitting Sight glass fitting for processes up to 10 bar (145 psi) Sight glass fitting for unpressurized processes Avoid material accumulation in front of the sight glass (risk of incorrect measure- ments)�...
Page 13
Solimotion FTR16 Mounting Mounting on process nozzle 0000000027 11 Mounting on process nozzle Mounting flange Process nozzle Plug Process insulation • Suitable mounting flange → 26 • Suitable plug → 28 • Suitable process nozzle of type FAR50 → 31 In case of risk of buildup ‣...
Page 14
Mounting Solimotion FTR16 Mounting on existing process nozzles 0000000029 13 Mounting with FAR51 insertion adapter Insertion adapter Customer-supplied gasket Customer-supplied installation material Process nozzle Suitable insertion adapter of type FAR51 → 31 Mounting with high pressure adapter ‣ Applications with process pressure up to 21 bar (305 psi) 0000000026 14...
Page 15
Solimotion FTR16 Mounting Mounting with high temperature adapter and extensions ‣ Applications with process temperature +60 to +450 °C (+140 to +842 °F) • Process pressure: 0.8 to 5.1 bar (12 ... 74 psi) absolute • Observe maximum temperature T on process connection of FTR16.
Electrical connection Solimotion FTR16 5�3 Post-installation check Is the device undamaged (visual inspection)? Does the device conform to the measuring point specifications? For example: • Process temperature • Process pressure • Ambient temperature range • Measuring range Are the measuring point number and labeling correct (visual inspection)? ...
Solimotion FTR16 Electrical connection • Maximum bulk material movement: The FTR16 keeps the electrical switch closed as long as there is bulk material move- ment. • Minimal or no bulk material movement The FTR16 keeps the electrical switch closed as long as there is minimal or no bulk material movement.
Operation options Solimotion FTR16 6�4 Post-connection check Is the device or cable undamaged (visual inspection)? Do the cables used comply with the requirements? Do the mounted cables have adequate strain relief? Are the connectors firmly tightened? ...
Solimotion FTR16 Operation options 7�1 Light signals (LEDs) Display Meaning Operation LED lights up: Device is ready for operation (supply voltage is pres- ent) LED flashes: Device is in parameterization mode (→ 20) Error LED lights up: Error/device failure (unrecoverable error).
Operation options Solimotion FTR16 7�3 On-site operation 0000000011 20 On-site operation Operation via operating magnet (→ 25) For operation, the operating magnet (north pole visible as shown) must be placed on the marked surfaces of the FTR16. The operating magnet is included in the scope of delivery.
Solimotion FTR16 Operation options • If there is no or only slight bulk material movement, the yellow LED (sensor status) report the lack of movement, and the white LED is off or flashing at a low frequency. If this is not the case, the process window must then be adjusted.
Diagnostics and troubleshooting Solimotion FTR16 7�4�5 Reset to factory settings In case of unknown settings or use in a new application, it is recommended to reset the FTR16 to its factory settings beforehand. Factory settings: • Process windows: 50 % •...
The measuring device must be returned if the wrong device has been ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in con- tact with medium.
Page 24
Accessories Solimotion FTR16 • Order number: - 71530949 (5 m (196.85 in)) - 71530950 (10 m (393.70 in) - 71530953 (20 m (787.40 in)) Connection cable with M12 angled socket (-/.) • Number of poles/cross section: 4 x 0.34 mm •...
Solimotion FTR16 Accessories 12�10 Insertion adapter • Type FAR51 (→ TI01368F) • Process nozzle - DN50 to DN100, PN16, Form A - NPS 2" to 4" 150 lbs, RF • Nozzle length: 100 to 300 mm (3.94 to 11.81 in) •...
Ex connection data See safety instructions (XA): All data relating to explosion protection are provided in separate Ex documentation and are available from the Downloads Area of the Endress+Hauser-website. The Ex documentation is supplied as standard with all Ex de- vices. Endress+Hauser...
Technical data Solimotion FTR16 13�3 Power supply 13�3�1 Supply voltage • U = 18 to 30 V DC • In accordance with IEC/EN61010 a suitable circuit breaker must be provided for the measuring device. • Voltage source: Non-hazardous contact voltage or Class 2 circuit (North America).
• 0.8 to 5.1 bar (12 ... 74 psi) absolute, with high temperature adapter • 0.5 to 21 bar (7 ... 305 psi) absolute, with high pressure adapter • Observe deviating pressure ranges for the accessories offered! 13�6 Additional technical data Latest technical information: Endress+Hauser website: www.endress.com → Downloads. Endress+Hauser...