Monitoring Devices And Indicators For Radiation Protection - Siemens LITHOSTAR Multiline Installation Instructions Manual

Hide thumbs Also See for LITHOSTAR Multiline:
Table of Contents

Advertisement

4 - 20

Monitoring devices and indicators for radiation protection

At the customer's request or when country-specific regulations require (e.g. DHHS), the
following monitoring devices and indicators can be connected in the POLYDOROS SX to
D160.X61 to 64 (1/Fig. 15):
Connection of the door contact for blocking radiation
After removing the associated jumper, connect the door contact
for AP1 to:
Connection of the indicator lamps for tube assembly selection
Attach an indicator lamp in the vicinity of the tube assembly (24 V ~, max. 5 W) so that it
is clearly visible from the control console.
Connect the lamp
for AP1 to:
LITHOSTAR Multiline
System
Connections with POLYDOROS SX 65
D160.X61.3 and D160.X61.4
24 V ∼ and D160.X61.6 (0 V
D160.X61.5
RXL2-120.812.01
Rev. 02
X41
X64
X63
X62
X61
Fig. 15