106 | DRY 3.4 | Advanced Operation (key-operator mode)
Performing quality control (QC) tests for mammography
application
The following procedures must be performed daily, weekly or annually as
indicated. When used in a mobile application it is required to perform QC
after every move.
The reason for performing quality control tests is to determine if any
significant image quality variation or deterioration has occurred which may
require corrective action. Comparing the results of the tests with the reference
values previously established does this.
This procedure allows the operator to take the necessary preventive actions
before any image quality loss can take place.
Topics:
Performing the daily QC test for mammography application
Performing the Weekly QC tests for mammography application
Performing the Annual QC tests for mammography application
Performing the daily QC test for mammography application
1. Turn on the printer and wait at least for 15 minutes.
2. Press the Key-operator key to enter the Key-operator mode.
The 'Key-operator main menu' screen appears:
3. Press the Down key eight times, followed by the Confirm key to select
'Quality Control'.
The 'Select input tray' screen appears:
2602A EN 20191018 0826
WARNING:
This test must be performed every day before any clinical film
can be processed.