Technical Specifications - ResMed VPAP Malibu User Manual

Hide thumbs Also See for VPAP Malibu:
Table of Contents

Advertisement

Technical Specifications

Performance
Operating pressure range: 3 to 25 cm H
Maximum single fault pressure: 40 cm H
Pressure measurement tolerance: ±0.5 cm H
Flow measurement tolerance: ±0.2 L/s
Dynamic pressure characteristics
VSet mode: 4 to 25 cm H
AutoSet Max pressure: 20 cm H
S mode: IPAP: 3 to 25 cm H
EPAP: 3 to 25 cm H
CPAP mode: 4 to 20 cm H
Dynamic flow characteristics
130 L/min at 3 to 25 cm H
Sound pressure level: <30 dB (tested in accordance with the requirements of
ISO 17510-1:2002)
Data storage: 365 days of usage, leak, pressure, AHI and AI data
Dimensions (L x W x H): 10.6" x 9.1" x 5.6"
Weight: 5.1lb
Power supply: Input range 100–240V, 50–60Hz, 40VA (typical power
consumption), < 100VA (maximum power consumption)
Housing construction: Flame retardant engineering thermoplastic
Environmental conditions
Operating temperature: 41
Operating humidity: 10–95% non-condensing
Storage and transport temperature: -4
Storage and transport humidity: 10–95% non-condensing
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial, and light industry
environments. For further details, see "Guidance and Manufacturer's
Declaration - Electromagnetic Emissions and Immunity" on page 37.
Air filter: Two-layered, powder-bonded, polyester non-woven fiber
Air tubing: Flexible plastic, 1 x 6'6" (2 m) or 1 x 9'10" (3 m) length
Air outlet: 22mm taper, compatible with EN 1281-1:1997 Anaesthetic &
Respiratory Equipment - Conical Connectors
IEC 60601-1 classification: Class II (double insulated), Type CF
Note: The manufacturer reserves the right to change these specifications
without notice.
2
O (measured at the end of standard 6'6" air tubing);
2
O; Pressure Support: 0 to 10 cm H
2
O (measured at the end of standard 6'6" air tubing);
2
O (measured at the end of standard 6'6" air tubing)
2
O (measured at the end of standard 6'6" air tubing)
2
O
2
o
o
F to 104
F (+5
o
F to 140
O
O
2
O + 4% of the measured reading
2
o
o
C to +40
C)
o
o
F (-20
C to +60
Technical Specifications
O
2
o
C)
35

Advertisement

Table of Contents
loading

Table of Contents