◆ The values are acceptable if both the
L value and S value are within ±1.0 of
the reference values indicated on the
checker cover. (The L and S values
indicate the hardware characteris-
tics.)
If the measured value exceeds ±1.0
of the reference value, clean both the
checker and probe, then take mea-
surement again.
If ±1.0 of the reference value is still
exceeded, contact the nearest KON-
ICA MINOLTA SENSING authorized
service facility.
7
Close the check cover.
8
Set the POWER switch to
OFF.
S value (measured value of
the short optical path)
17