Device Messages - ResMed AirCurve 10 VAUTO Clinical Manual

Hide thumbs Also See for AirCurve 10 VAUTO:
Table of Contents

Advertisement

Problem/possible cause
Water tub is leaking
Water tub may not be assembled correctly.
Water tub may be damaged or cracked.
The patient's therapy data has not been transmitted
Wireless coverage may be poor.
The No wireless connection icon
top right of the screen. No wireless network available.
Device may be in Airplane Mode.
SmartStart is enabled, but the device does not automatically start when the patient breathes into their mask
Breath is not deep enough to trigger SmartStart.
There is excessive leak.
SmartStart is enabled, but the device does not automatically stop when the patient removes their mask
Incompatible mask being used.

Device messages

Device message/possible cause
High leak detected, check your water tub, tub seal or side cover
Water tub may not be inserted properly.
Water tub seal may not be inserted properly.
High leak detected, connect your tubing
Air tubing may not be connected properly.
Solution
Check for damage and reassemble the water tub correctly.
Replace the water tub.
Advise the patient to place the device where there is
coverage (ie, on their bedside table, not in a drawer or on the
floor).
The Wireless signal strength icon
coverage when all bars are displayed, and poor coverage
when fewer bars are displayed.
Advise the patient that therapy data can be sent using the
is displayed on the
SD Card.
Turn off Airplane Mode, for instructions see the User Guide.
To start therapy, take a deep breath in and out through the
mask, before breathing normally.
Press Start.
Adjust the mask and headgear.
Air tubing may not be connected properly. Connect firmly at
both ends.
Only use equipment recommended by ResMed.
Contact ResMed or see www.resmed.com for more
information.
If the patient is using a nasal pillows mask with set pressure
less than 7 cm H
should be disabled.
Solution
Make sure the water tub is correctly inserted.
Open the water tub and make sure that the seal is correctly
inserted.
Make sure the air tubing is firmly connected at both ends.
indicates good
O (7 hPa), SmartStart will not work and
2
English
33