Linefeed; Ieee-488 Parameters; Ieee-488 Address - Fluke Hart Scientific 7100 User Manual

Calibration bath
Table of Contents

Advertisement

dUPL
Press "SET" to access the mode setting.
dUP=FULL
The mode may be changed using "UP" or "DOWN" and pressing "SET".
dUP=HALF
9.12.4

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
Press "SET" to access the linefeed parameter.
LF= On
The mode may be changed using "UP" or "DOWN" and pressing "SET".
LF= OFF
9.13

IEEE-488 Parameters

Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case
the user may set the interface address within the IEEE-488 parameter menu.
This menu does not appear on baths not fitted with the interface. The menu is
indicated by,
Press "SET" to enter the menu.
9.13.1

IEEE-488 Address

The IEEE-488 interface must be configured to use the same address as the ex-
ternal communicating device. The address is indicated by,
AddrESS
Press "SET" to access the address setting.
Serial duplex mode parameter
Current duplex mode setting
New duplex mode setting
Serial linefeed parameter
LF
Current linefeed setting
New linefeed setting
IEEE-488 parameters menu
IEEE
IEEE-488 interface address
9 Controller Operation
IEEE-488 Parameters
49