7100 Calibration Bath
User's Guide
4.4
12
TOP VIEW OF COMPRESSOR
Adjust
BACK OF BATH
Figure 2 Semi-hermetic Compressor
tion5
for detailed instructions for proper installation of the bath. Be sure to
place the bath in a safe, clean and level location.
Fill the bath tank with an appropriate fluid. For operation at bath temperatures
above 0°C, clean distilled water works well. For lower temperatures, fluids
such as methanol and Halocarbon may be used. Carefully pour the fluid into
the bath tank through the access hole above the tank avoiding spilling any fluid.
The fluid must not exceed a height of 1/2 inch below the top of the tank.
The control probe must be inserted through the lid into the bath and plugged
into the socket at the back of the bath. The bath stirrer and haeater must be
plugged in. Connectors are different to prevent improper connections.

Power

Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability: 230 VAC (±10%), 60 Hz, 15 amp [50 Hz optional]. Set
the "HEATER" switch on the front panel to position "LOW" and turn the bath
on using the front panel "POWER" switch. The bath will turn on and begin to
heat or cool to reach the previously programmed temperature set-point. The
front panel LED display will indicate the actual bath temperature.
Adjust
HARDWARE ADJUSTED
FOR SHIPPING
HARDWARE ADJUSTED
FOR OPERATION