Replace the pretreatment and polisher modules as follows:
按照下列方法更换预处理柱以及抛光柱
• • • •
First insert the module in the lower connector, then the top connector.
• • • •
Gently press the module into the connections until you feel them lock into place.
• • • •
Push the module downwards onto the lock.
• • • •
Check that the new module is correctly and firmly installed.
Close the dispenser again.
Turn on the water supply and plug in power cord.
Switch on the device using the MENU key.
Use the dispenser and the sterile filter to purge the system:
unscrew the purge screw on the sterile filter and open the dispenser until the water flows out
without bubbles (Warning! Wear protective goggles as water will come out).
Check all connections for leaks.
Close the front panel.
The system is ready for use.
先将过滤柱的下端对准连接器,然后上端;
轻轻用力将过滤柱压入连接器,直到你感觉到它们已经就位;
向下将过滤柱推进锁槽内;
检查新的过滤柱已经安装好;
LaboStar 7 TWF DI and UV
Ultra pure water system
Catalog no. 2231 & 2232