7.2

Module replacement

Switch the device off using the MENU key and unplug power cord.
Turn off the feed water.
Open the dispenser and wait for system pressure to dissipate.
Open the front panel.
Push module upwards off the lock and remove from the connections by pulling outward.
使用"MENU"键关掉纯水机并将电源线拔掉;
关闭进水阀门;
打开供水龙头直到系统管路中没有压力;
打开前端门板;
将过滤柱向上推离锁槽,然后向外把过滤柱从连接器中拔出来。
IMPORTANT!
重 重 重 重 要 要 要 要
更 更 更 更 换 换 换 换 过 过 过 过 滤 滤 滤 滤 柱 柱 柱 柱
Connections
连接器
Polisher module
抛光住
锁槽
Lock
When removing the modules, take the top end out of the bracket first, then
the bottom end. When replacing the modules, be careful not to damage the
plastic pins in the connections.
取出过滤柱时,先将上端脱开,然后下端脱开。当更换过滤柱时,请注意不
要将连接器中的塑料插脚弄断。
LaboStar 7 TWF DI and UV
Ultra pure water system
Catalog no. 2231 & 2232