c.
Call the NI 9220 get C Series adjustment points function to obtain an array of
recommended calibration voltages for the NI 9220.
d.
Set the calibrator to a reference value determined by the array of recommended
calibration voltages.
e.
Set the calibrator to Operate mode (OPR).
f.
Call and configure the NI 9220 adjustment function according to Table 5.
Physical Channel
cDAQMod8/aix
g.
Set the calibrator to Standby mode (STBY).
h.
Repeat steps d through g for each calibration voltage in the array.
i.
Close the calibration session.
4.
Disconnect the NI 9220 from the calibrator.
5.
Repeat steps 1 through 4 for each channel on the NI 9220.

EEPROM Update

When an adjustment procedure is completed, the NI 9220 internal calibration memory
(EEPROM) is immediately updated.
If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments by initializing an external calibration,
setting the C Series calibration temperature, and closing the external calibration.

Re-Verification

Repeat the
Verification
If any test fails Re-Verification after performing an adjustment, verify that
Note
you have met the
Services
6 | ni.com | NI 9220 Calibration Procedure
Table 5. Adjustment Configuration
The reference value from the array of adjustment points
section to determine the As-Left status of the device.
Test Conditions
section for assistance in returning the device to NI.
Reference Value
before returning your device to NI. Refer to