Summary of Contents for National Instruments 9224

  • Page 1 NI-9224...
  • Page 2: Table Of Contents

    EEPROM Update........................6 Reverification..........................6 Accuracy under Calibration Conditions..................6 Worldwide Support and Services....................6 Software Calibrating the NI 9224 requires the installation of NI-DAQmx 18.1 or later on the calibration system. You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports LabVIEW, ™ ™...
  • Page 3: Test Equipment

    Table 1. Recommended Equipment Equipment Recommended Model Requirements Use a high-precision voltage source with gain error of Calibrator Fluke 5502A <500 ppm and offset error <50 μV. Chassis NI cDAQ-9178 — 2 | ni.com | NI 9224 Calibration Procedure...
  • Page 4: Test Conditions

    Verification The following performance verification procedure describes the sequence of operation and provides test points required to verify the NI 9224. The verification procedure assumes that adequate traceable uncertainties are available for the calibration references. Accuracy Verification Complete the following procedure to determine the As-Found status of the NI 9224.
  • Page 5 If the value is within these limits, the device passes the test. Repeat steps 3 through 7 for each test point in the NI 9224 Verification Test Limits table. Disconnect the calibrator from the device.
  • Page 6: Adjustment

    Initialize a calibration session on the NI 9224. The default password is Input the external temperature in degrees Celsius. Call the NI 9224 get C Series adjustment points function to obtain an array of recommended calibration voltages for the NI 9224.
  • Page 7: Eeprom Update

    EEPROM Update When an adjustment procedure is completed, the NI 9224 internal calibration memory (EEPROM) is immediately updated. If you do not want to perform an adjustment, you can update the calibration date and onboard calibration temperature without making any adjustments by initializing an external calibration, setting the C Series calibration temperature, and closing the external calibration.
  • Page 8 1 866 ASK MYNI (275 6964). For support outside the United States, visit the Worldwide Offices section of ni.com/niglobal access the branch office websites, which provide up-to-date contact information. NI 9224 Calibration Procedure | © National Instruments | 7...
  • Page 9 For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your patents.txt ni.com/patents software, the file on your media, or the National Instruments Patent Notice at . You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer ni.com/legal/export-compliance...