Pentair HAFFMANS GMT Instruction Manual

Pentair HAFFMANS GMT Instruction Manual

Analogue co2 gehaltemeter

Advertisement

HAFFMANS
HAFFMANS
ANALOGUE CO
INSTRUCTION MANUAL
GEHALTEMETER
2
TYPE GMT
W W W.HAFFMANS.NL
W W W.HAFFMANS.NL

Advertisement

Table of Contents
loading

Summary of Contents for Pentair HAFFMANS GMT

  • Page 1 HAFFMANS HAFFMANS INSTRUCTION MANUAL ANALOGUE CO GEHALTEMETER TYPE GMT W W W.HAFFMANS.NL W W W.HAFFMANS.NL...
  • Page 3 Serial number: from GMT-7828 Essential reading! Haffmans BV is not responsible for any damage caused by the owner’s failure to follow the instructions given in this manual. Haffmans BV is not responsible for print or translation errors in this edition of the manual. The specifications and design of this product are subject of change without notice.
  • Page 5: Table Of Contents

    HAFFMANS  GMT   Table of contents   Technical specifications ....................7 Introduction........................9 Intended use ......................9 Safety ........................9 General product information ..................11 Installation........................13 Unpacking ......................13 Installing the battery .................... 13 Operation........................15 Calibration ........................17 Determining the accuracy of the pressure gauge ..........17 Adjusting the handle....................
  • Page 6 HAFFMANS  GMT     Page 6 of 37...
  • Page 7: Technical Specifications

    HAFFMANS  GMT   TECHNICAL SPECIFICATIONS   Measuring range Thermometer -2 / +20 ºC Manometer 0 / 2,5 0,20 / 1,20 gew. % Pressure range Accuracy Manometer Thermometer ºC Dimensions Height without grip Width Depth Weight Battery type 9V Alkaline Scale bar, ºC, weight. % Protection class IP 64 Page 7 of 37...
  • Page 8 HAFFMANS  GMT     Page 8 of 37...
  • Page 9: Introduction

    HAFFMANS  GMT   INTRODUCTION   Intended use The CO Gehaltemeter is only intended to be used to determine the gas equilibrium pressure and the temperature of a liquid when calculating the CO content. The instructions in this manual refer to the position numbers given in the diagram in appendix Safety This section gives the most important instructions for the safe use of the CO Gehaltemeter.
  • Page 10 HAFFMANS  GMT     Page 10 of 37...
  • Page 11: General Product Information

    HAFFMANS  GMT   GENERAL PRODUCT INFORMATION   The CO Gehaltemeter is based on Henry’s law, which applies for an equilibrium state between a gas and a liquid in a confined space. This law states that the quantity of gas dissolved in a liquid at a certain temperature is proportional to the partial pressure of the gas above the liquid.
  • Page 12 HAFFMANS  GMT     Page 12 of 37...
  • Page 13: Installation

    HAFFMANS  GMT   INSTALLATION   Unpacking Before despatch the GMT was checked and tested by Haffmans BV Make sure that the content of the delivery is complete and undamaged. If the delivery is incomplete or damaged, please contact Haffmans BV or the representative/agent designated by Haffmans B.V, immediately (appendix IV).
  • Page 14 HAFFMANS  GMT     Page 14 of 37...
  • Page 15: Operation

    HAFFMANS  GMT   OPERATION   The CO Gehaltemeter is suitable for measuring pressures up to 5 bar. It is extremely dangerous to allow the pressure in the Gehaltemeter to increase to more than 5 bar, because the Gehaltemeter may explode. Warning The pressure in the instrument must not exceed the range of the pressure gauge whilst taking samples (otherwise the pressure gauge will break).
  • Page 16 HAFFMANS  GMT   Example:   Information: Temperature (T) = + 1.5 Pressure (P) = 0.7 bar The slide rule gives a CO value equal to 0.51 percent of the weight (weight %). Page 16 of 37...
  • Page 17: Calibration

    HAFFMANS  GMT   CALIBRATION   Determining the accuracy of the pressure gauge The CO Gehaltemeter’s pressure gauge is accurate to 1% of its total scale. The pressure gauge can be calibrated using a Haffmans Gauge Calibration Device (GCD). Method: Attention! The pressure gauge is calibrated for the most common pressures, i.e.
  • Page 18: Adjusting The Handle

    HAFFMANS  GMT   Adjusting the handle   The handle is set so that the electrical connection is made before the measurement is started, just before the handle is fully depressed. The handle does not normally need to be adjusted, but can be adjusted if necessary. Method: Attention! Do not unscrew the adjusting screw too far, because the handle...
  • Page 19: Maintenance And Cleaning

    HAFFMANS  GMT   MAINTENANCE AND CLEANING   Daily The CO Gehaltemeter must be thoroughly rinsed with cold water after all the measurements have been taken. Method: Attention! The temperature in the room must not fall below 0 1. Connect the supply hose to the water tap, 2.
  • Page 20: Monthly

    HAFFMANS  GMT   Monthly   Cleaning the burette and plunger Burette (pos. 111) Method: 1. Unscrew the burette’s attachment screws (pos. 73, side of the GMT), 2. Turn the safety plates (pos. 84) upwards, 3. Slide the burette (pos. 111) sideways out of the CO Gehaltemeter, 4.
  • Page 21: Trouble Shooting And Repair

    HAFFMANS  GMT   TROUBLE SHOOTING AND REPAIR   Trouble shooting Fault Possible cause Solution Page The plunger leaks. The O-rings are Disassemble the plunger and damaged. clean or replace the damaged O- rings. The handle has been Adjust the handle. set incorrectly. The plunger does not The plunger and/or the Disassemble and clean the...
  • Page 22: Repair

    HAFFMANS  GMT   Repair   8.2.1 Replacing the pressure gauge Method: 1. Unscrew the pressure gauge’s (pos. 67) cover. 2. Using a pressure gauge puller (included in the service set), pull the pressure gauge out of the instrument whilst rotating it. 3.
  • Page 23 HAFFMANS  GMT   8.2.3 Replacing the thermometer   Method: 1. Remove the burette (pos. 111) according to the instructions given on page 14, 2. Fit the thin wire (included in the service set) behind the thermometer are carefully tilt it, 3. Use your thumb to exert a small counter pressure on the top of the thermometer and pull it, together with the wire, out of the instrument, 4.
  • Page 24 HAFFMANS  GMT     Page 24 of 37...
  • Page 25: Disposal And The Environment

    HAFFMANS  GMT   DISPOSAL AND THE ENVIRONMENT   If the CO Gehaltemeter has to be disposed off, the various components must be disposed of separately according to the environmental regulations of the country concerned. The most common method is to separate the stainless steel and plastic components. The battery and thermometer must also be disposed of separately.
  • Page 26 HAFFMANS  GMT     Page 26 of 37...
  • Page 27: Appendix I: Customer Service

    HAFFMANS  GMT   APPENDIX I: CUSTOMER SERVICE   Warning Repairs to the electrical parts of the CO Gehaltemeter should always be carried out by a Haffmans service engineer. The trouble shooting chart list a number of potential minor problems and indicates how the customer can solve these problems.
  • Page 28 HAFFMANS  GMT     Page 28 of 37...
  • Page 29: Appendix Ii: Terms Of Guarantee

    HAFFMANS  GMT   APPENDIX II: TERMS OF GUARANTEE   This Haffmans guarantee is in addition to all the rights accruing to the buyer from the sales contract between buyer and supplier and the relevant legislation. Haffmans BV guarantees that the equipment supplied is free from material and manufacturing defects, for normal use and maintenance.
  • Page 30 HAFFMANS  GMT     Page 30 of 37...
  • Page 31: Appendix Iii: Ec Declaration Of Conformity

    HAFFMANS  GMT   APPENDIX III: EC DECLARATION OF CONFORMITY   TRANSLATION OF THE ORIGINAL EC DECLARATION OF CONFORMITY EC DECLARATION OF CONFORMITY HAFFMANS B.V. Marinus Dammeweg 30 5928 PW Venlo, Holland declare entirely under our own responsibility, that the product -MEASURING, TYPE GMT upon this statement has reference to, is in conformity with the norms EN 61000-6-1, EN 61000-6-3 according to the demands of:...
  • Page 32 HAFFMANS  GMT     Page 32 of 37...
  • Page 33: Appendix Iv: Spare Parts

    HAFFMANS  GMT   APPENDIX IV: SPARE PARTS   When you order spare parts please include all the following information with your order: • Serial number of the CO Gehaltemeter. • Position number (on the parts list below). • Description. • Article Number. Gehaltemeter Drawing number GMT-00-1007 Pos.no.
  • Page 34 HAFFMANS  GMT     Pos.no. Description Art. no. Screw M4x10 000.770 O-ring 29,82x2,62 000.772 Screw cap for battery housing 000.773 Screw M4x 15 000.774 O-ring 40,95x2,62 000.775 Battery 9V 108.046 Safety plate 000.777 O-ring 15,55x2,62 000.778 Stud 000.780 000.781 000.781 Actuating rod 000.783 93-1 Adjusting screw...
  • Page 35 HAFFMANS  GMT     Pos.no. Description Art. no. Manometer holder with O-ring packing (part no. 5) 000.804 Cupper ring 000.805 Plunger 000.332 Hinge block 000.807 Thermometer –2/+20 degrees Celsius 000.720 Burette 000.824 Electrical unit 000.826 Carrying handle (standard) complete 003.909 Hose adapter 100.019 O-ring 7.65x1.78 001.630...
  • Page 36 HAFFMANS  GMT     Page 36 of 37...
  • Page 37: Appendix V: Drawing

    HAFFMANS  GMT   APPENDIX V: DRAWING   Description Drawing number Composition GMT GMT-00-1007 Page 37 of 37...