  • Page 3 c-DGM from software version: 2.04-11 i-DGM from software version: 1.20-06 Essential reading! Haffmans B.V. is not liable for damage that results from failure to follow the instructions in this manual. Furthermore, Haffmans B.V. is not liable for writing/typing errors in this publication. The specifications and design of the instrument may be changed without advance notice.
  • Page 5: Table Of Contents

    , &&D E^ 'D 'D Table of contents Technical specifications Introduction Intended use Safety Symbols Precautionary measures and safety instructions General product information General description Measurement principle measurement measurement *TPO measurement Installation Contents of the delivery Preparation Units Time settings 5.4.1 Date 5.4.2...
  • Page 6 , &&D E^ 'D 'D 6.2.4 Continuous *O measurement 6.2.5 *Combined CO / TPO measurement 6.2.6 *TPO measurement Reading the measurement results Erasing measurement results Product choice Charging the battery Calibration and compensation Calibration of the temperature sensor Calibration of the pressure 7.2.1 Calibration instructions Compensation...
  • Page 7: Technical Specifications

    , &&D E^ 'D 'D Technical specifications Measurement range : Default : [g/l] : 2.00 to 10.00 : Available : [Vol.] : 1.00 to 5.00 : [% b.w.] : 0.200 to 1.000 : [kPa] : 21.6 to 510.9 : [kg/cm : 0.02 to 5.21 : Default : [ppb w/w]...
  • Page 8 , &&D E^ 'D 'D * at 20 ° C : *O dissolved :[ppb w/w] : ± (1 ppb+2% LHO (DO) of the measured value) : [µg/l] : ± (1 µg/l +2% otmv) : [ppm w/w] : ± (0,001 ppm+2% otmv) : [mg/l] : ±...
  • Page 9 , &&D E^ 'D 'D Duration of measurement measurement : approx. 55 seconds *Combined CO : approx. 120 seconds measurement measurement : approx. 60 seconds : meas. every 10 to 999 seconds continuous measurement *Combined CO /TPO : approx. 130 seconds measurement *TPO meting : approx.
  • Page 10 , &&D E^ 'D 'D Page 10 of 78 c-DGM/i-DGM...
  • Page 11: Introduction

    , &&D E^ 'D 'D Introduction Intended use The digital content meter (c-DGM and i-DGM) is a portable, chargeable instrument to determine the dissolved O and CO content of carbonated drinks and the O content in gases. The content meter is filled with carbonated drink originating from, for example: the production process (tanks and piping) kegs filled bottles and cans...
  • Page 12 , &&D E^ 'D 'D Page 12 of 78 c-DGM/i-DGM...
  • Page 13: Safety

    , &&D E^ 'D 'D Safety Symbols The symbols Attention!, Warning and Danger have the meaning indicated below in this manual. Attention! Instructions for correct and effective use. Warning Incorrect or careless use may seriously damage the instrument. Incorrect or careless use may result in DANGER to the user and the surroundings.
  • Page 14 , &&D E^ 'D 'D Page 14 of 78 c-DGM/i-DGM...
  • Page 15: General Product Information

    , &&D E^ 'D 'D General product information General description The DGM has two types of controls. 1. Push buttons on the top of the instrument. Four mechanical component, consisting of: the handle (pos. 20), the throttle valve (pos. 12), and the two hose connectors for the sampling hoses.
  • Page 16: Measurement Principle

    , &&D E^ 'D 'D Measurement principle measurement measurement is based on measuring the luminescence of an oxygen-sensitive layer. The fluorescence changes depending on the partial oxygen pressure. The quantity of dissolved gas in the liquid is calculated by using the partial oxygen pressure and temperature. measurement The measurement of CO is based on Henry's law, which describes the equilibrium between a...
  • Page 17: O 2 Measurement

    , &&D E^ 'D 'D measurement The oxygen sensor (pos. 23) is located behind the inlet (pos. 17) of the DGM and in front of the inlet to the CO measurement chamber (pos. 02). There are two different types of oxygen sensor available for the c-DGM. The LHO sensor is suitable for measuring very accurately at low oxygen concentrations, the WLO sensor is suitable for measuring for higher oxygen concentration over a wider measuring range.
  • Page 18: Tpo Measurement

    , &&D E^ 'D 'D *TPO measurement When the c-DGM is used in combination with an ISD sampling device, the TPO value can be determined in a sealed product packet. The traditional Uhlig's method, whereby the DO value is multiplied by a Z factor (see Appendix I) is used to calculate the TPO value. The use of Uhlig's method is based on an oxygen balance between the oxygen dissolved in the liquid and the oxygen in the headspace volume.
  • Page 19: Installation

    , &&D E^ 'D 'D Installation Contents of the delivery The o-DGM has been calibrated, checked and tested by Haffmans BV before delivery so that it is ready for immediate use. It is not necessary to calibrate the instrument. Make sure the delivery is complete and undamaged. If the delivery is incomplete or damaged, immediately notify Haffmans B.V.
  • Page 20: Preparation

    , &&D E^ 'D 'D Preparation Make sure that the DGM is not connected to a sampling point that produces pressure or pressure impulses exceeding 10 bar. Connect the hose set. Method: 1. Lubricate the O-rings (pos. 18) of the hose socket (pos. 17) with a small amount of silicon grease (from the service set).
  • Page 21: Time Settings

    , &&D E^ 'D 'D Time settings You can change the various time settings by means of the “Date/Time/Display” menu of the DGM. The following steps take you to the “Date/Time/Display” menu: 1. Press “Enter” until the Haffmans logo appears. 2.
  • Page 22: Check Electrolysis Current

    , &&D E^ 'D 'D 5.4.5 Check electrolysis current An electrolysis current check is performed to measure the CO . The warning “No current strength for electrolysis measured. Measurement aborted.” appears in the case of low conductivity. If liquids with low conductivity are measured, it is possible to switch off the electrolysis current test by setting the option “Check E-Current”...
  • Page 23: Product

    , &&D E^ 'D 'D 5.5.1 Product You can now browse through the 10 products and press “Enter” to select a product. While browsing through the products, the settings for the currently selected product are displayed. You can enter a product name consisting of a maximum of 10 characters. The cursor moves to the next character each time “Enter”...
  • Page 24: Electrolysis Time

    , &&D E^ 'D 'D 5.5.3 Electrolysis time You can change the duration of the electrolysis time with the arrow buttons. When you press “Enter”, the cursor jumps to the following digit; if you press “Enter” again, you return to the “CO formula/ prod.def.”...
  • Page 25: Co 2 Stabilization Time

    , &&D E^ 'D 'D 5.5.4 CO stabilization time You can change the duration of the electrolysis time with the arrow buttons. When you press “Enter”, the cursor jumps to the following digit; if you press “Enter” again, you return to the “CO formula/ Prod.Def.”...
  • Page 26: Language

    , &&D E^ 'D 'D Language Change the language setting as follows: 1. Press “Enter” until the Haffmans logo appears. 2. Press “Enter” again. This takes you to the main menu. 3. Use the arrow buttons to select the “Configuration” menu and then press “Enter” to open the menu.
  • Page 27: Password

    , &&D E^ 'D 'D Activate the ID system for the “User-ID” or the “Place-ID” in the following manner: 1. Select “Place ID” or “User ID” with the arrow buttons and press “Enter”. The ID is set to “Absent”. 2. Press one of the arrow buttons. The ID is set to “Present”.
  • Page 28 , &&D E^ 'D 'D First enter the old password, and then enter a new 4-digit password. Use the arrow buttons to change the numbers and press “Enter” to go to the next digit. Attention! If you change the password, remember it or write it in the space below.
  • Page 29: Operation

    , &&D E^ 'D 'D Operation General The c-DGM has 6 different measuring programs to choose from: Combined CO and O measurement measurement measurement Continuous O measurement Combined CO and TPO measurement TPO measurement The desired program can be selected in the “Measurement choice” menu (see 6.2). The “Measurement choice”...
  • Page 30 , &&D E^ 'D 'D Method: 1. Connect the DGM sampling hose to the sampling point. Place the end of the outlet hose (DGM) in a drain and secure it in place. Attention! When samples are under low pressure, the instrument must be placed as low as possible.
  • Page 31 , &&D E^ 'D 'D If the RFID system (optional) is active, you will be asked to identify yourself (User ID). To do so, hold the operator key against the front of the instrument (see fig. 6.1). Next, give the measurement location by holding the front of the DGM against the location key.
  • Page 32: Measurement Choice

    , &&D E^ 'D 'D *Measurement choice For the c-DGM you can select one of the following measurement programs from the “Measurement choice” menu: 1) Combined CO measurement 2) Only CO measurement 3) Only O measurement 4) Continuous O measurement 5) Combined CO / TPO measurement 6) TPO measurement...
  • Page 33 , &&D E^ 'D 'D 6.2.1 *Combined CO measurement During the combined CO measurement, the c-DGM carries out the following process steps: oxygen measurement, display of the measured O value, electrolysis, pause for stabilization of the CO equilibrium pressure and temperature, display of the measured temperature, display of the measured pressure, and display of the measured CO content.
  • Page 34 , &&D E^ 'D 'D The figures below show the sequence of displays for a measurement : T: 5.2 P: 1:03 AM O: 77 Fig. 6.5 Oxygen content display BUSY MEASURING... PROGRESS: T: 5.2 P: 1:03 AM O: 77 Fig. 6.6: Display during electrolysis Tel.
  • Page 35: Measure Only

    , &&D E^ 'D 'D 6.2.2 Measure only CO The measurement proceeds according to the same process (see Fig. 6.4) as the combined CO measurement, except that the oxygen measurement at the beginning is skipped. The process steps related to the O measurement are eliminated.
  • Page 36 , &&D E^ 'D 'D 6.2.4.2 *O continuous measurement in gas With this measurement facility, the oxygen content is measured continuously while the gas to be measured flows through the c-DGM. This setting is typically used when filling systems with CO where for example the switching point from air to CO must be found.
  • Page 37: Combined Co 2 / Tpo Measurement

    , &&D E^ 'D 'D 6.2.5 *Combined CO / TPO measurement With this measurement option the CO and O proceeds in the same manner as in the combined CO measurement. However, the total oxygen content of a package is calculated based on the dissolved O value measured.
  • Page 38: Tpo Measurement

    , &&D E^ 'D 'D Inspection template An inspection template can be used to measure the headspace volume separately for every sample. The liquid volume can then be read on it. If the package volume is known, the headspace volume can also be calculated in this way. However, this method only works for bottles.
  • Page 39: Reading The Measurement Results

    , &&D E^ 'D 'D Reading the measurement results Procedure: 1. Press “Enter” for approximately 3 seconds. The input menu is shown. 2. Press the “Enter” button. The main menu is displayed. 3. Use the Down arrow button to select “Logbook” in the main menu and press “Enter”. 4.
  • Page 40: Charging The Battery

    , &&D E^ 'D 'D Charging the battery The battery level bar is divided into 5 blocks and is visible when the battery is fully charged. When the battery level bar shows less than 20% remaining charge, the battery pack must be recharged. In this condition, the battery still has enough charge to carry out several measurements before the display switches off and no additional measurements are possible.
  • Page 41 , &&D E^ 'D 'D Quick-charger (optional) A quick-charger is available as an optional extra (see appendix VI) and can be used to charge the battery pack separately. Procedure: 1) Use the tools from the service set to remove the battery pack's two attachment screws at the rear of the instrument.
  • Page 42 , &&D E^ 'D 'D Page 42 of 78 c-DGM/i-DGM...
  • Page 43: Calibration And Compensation

    , &&D E^ 'D 'D Calibration and compensation Calibration of the temperature sensor The temperature sensor is an accurate PT-100 that is associated with tight tolerances and therefore does not require calibration. A check can be performed by initiating a flow of liquid through the measurement chamber and comparing the temperature displayed on the DGM with a reading from a precision thermometer placed in the outlet of the DGM.
  • Page 44: Calibration Instructions

    , &&D E^ 'D 'D 7.2.1 Calibration instructions To calibrate the pressure sensor, proceed as follows. You can accept the value with the “Enter” button and change the value with the Up and Down arrow buttons. Make sure you have selected the correct units! In order to perform the calibration correctly and accurately, you must know the local atmospheric pressure.
  • Page 45 , &&D E^ 'D 'D 5. Retract the plunger (sampling position). Press the “Enter” button once to enter the atmospheric pressure. Calibration pressure The cursor jumps to the value after P-Atm : xx.xx bar parameter “P-Atm”. P-Low 0.0 bar You can accept the value with the “Enter” button and P-high : xx.x bar change the value with the Up and Down arrow...
  • Page 46: Compensation

    , &&D E^ 'D 'D Compensation 7.3.1 General There may be a discrepancy between the CO content as measured by the Digital Content Meter (DGM) and that of other measurement systems. This discrepancy can be corrected with compensation factor “A”, as described in the compensation instructions. The new code “A” value can be calculated as follows: Calibrated CO value...
  • Page 47: Calibration O 2 Sensor

    , &&D E^ 'D 'D *Calibration O sensor Because the oxygen sensor only undergoes a very slight amount of wear as a result of exposure, regular calibration is not required. It has been experimentally determined that after 1 million exposures (i.e. 1 million O measurements) the O deviation from the zero point is merely 5 ppb and at higher O...
  • Page 48: One Point Calibration

    , &&D E^ 'D 'D 7.4.2 *One point calibration LHO sensor For most applications in the low oxygen measuring range a calibration of the zero point is sufficient. To do so follow the instructions as given for the calibration with two gasses. The atmospheric pressure and the concentration of the high calibration do not have to be entered again.
  • Page 49 , &&D E^ 'D 'D Attention! Use only low gas flows to flush the c-DGM. Use a reducing valve while flushing the c-DGM with pressurized air and adjust the flow by means of a flow meter to 30 l/hour. Attention! For an accurate O calibration or gas inspection, the c-DGM must be at the same temperature as the gas.
  • Page 50 , &&D E^ 'D 'D The option “P-ATM” appears only if the c-DGM CALIBRATION is set to the oxygen units % or %a.s. In this SENSOR TYPE SENSOR TYPE SENSOR TYPE SENSOR TYPE :LHO case the set atmospheric pressure is taken into MEDIUM :GAS account while displaying the oxygen value.
  • Page 51 , &&D E^ 'D 'D 11. Press the “Enter” button to set the O sensor constants CALIBRATION automatically. P-ATM. :1.013 BAR ZERO PNT : 0.000 % HIGH : 4.010 % O O O O SENSOR CONSTANTS SENSOR CONSTANTS SENSOR CONSTANTS SENSOR CONSTANTS 2 2 2 2 STORE CALIBRATION...
  • Page 52 , &&D E^ 'D 'D Page 52 of 78 c-DGM/i-DGM...
  • Page 53: Cleaning And Maintenance

    , &&D E^ 'D 'D Cleaning and maintenance Daily cleaning The DGM must be rinsed with cold water every day after use. Connect the inlet hose to a cold water tap and rinse the instrument with the throttle valve (pos. 12) open and the handle (pos. 20) must be retracted to the open position.
  • Page 54 , &&D E^ 'D 'D If there are salt deposits in the instrument (beer stone), the instrument must also be cleaned at room temperature, as described above, by using an acidic cleaning agent with a nitric acid base (~ 0.5%) or a phosphoric acid base (~ 2%). Once completed, the instrument must be well rinsed with cold water.
  • Page 55 , &&D E^ 'D 'D 7. Replace damaged O-rings. Make sure the sealing surface is not damaged! Only use O-rings from the service set. 8. Apply a thin layer of silicone grease (from the service set) to the plunger and the O-rings.
  • Page 56 , &&D E^ 'D 'D Page 56 of 78 c-DGM/i-DGM...
  • Page 57: Faults And Repair

    , &&D E^ 'D 'D Faults and repair Fault list Problem Cause Solution Too many bubbles in the (a) defective O-ring(s) on Remove the plunger and replace the measurement chamber the plunger defective O-rings. during sampling Nothing appears on the The battery is discharged Recharge the battery pack.
  • Page 58: Replacing The Battery Pack

    , &&D E^ 'D 'D *Calibrate or replace O The maximum number of Calibrate or replace oxygen sensor. sensor exposures has been reached and the maximum symbol oxygen sensor deviation at on the display the zero point is 5 ppb. sensor values unstable During calibration, the set Recalibrate this calibration point;...
  • Page 59: Sensor

    , &&D E^ 'D 'D *Replacing the O sensor The replacement set for the oxygen sensor includes a new sensor, O-rings and a special screwdriver. Procedure: Rinse the c-DGM via the outlet with gas to dry the measuring chamber (pos. 2) and the plunger block (pos. 26). Warning If there is still liquid in the plunger block, this liquid may enter the electronics when the plunger block is removed.
  • Page 60 , &&D E^ 'D 'D Push the new oxygen sensor (pos. 23) into the plunger block from the top until it is fully seated. Replace the O-rings (pos. 28 and 29) in the plunger block with new O-rings from the oxygen sensor set and apply a light coat of silicone grease to the O- rings.
  • Page 61: Decommissioning, Storage And Transport

    , &&D E^ 'D 'D Decommissioning, storage and transport 10.1 Decommissioning When a DGM must be disposed of, the components must be sent to an appropriate authority for recycling; alternatively, the entire unit can be returned to Haffmans B.V. The following is a list of the important elements contained in the main components: PCB housing : metal plate, electronics and sensors, ABS...
  • Page 62 , &&D E^ 'D 'D Page 62 of 78 c-DGM/i-DGM...
  • Page 63: Appendix I Formulas

    , &&D E^ 'D 'D Appendix I Formulas Haffmans CO formula Attention! The partial pressure of foreign gases is not included in the calculation.         ° gewichts × × = conversion factor/compensation factor (1*) = gas equilibrium pressure (partial) = atmospheric pressure in bar = product-dependent factor (-10.74*)
  • Page 64 , &&D E^ 'D 'D Conversion of the pressure and temperature units: Pressure P(bar) P(psi)/14.5 P(bar) P(kg/cm²)/1.02 Temperature T(° C) (T(° F)-32)×5/9 Rule of thumb for the conversion of the CO unit (g/L) = 2 * CO (volume) (g/L) = 10 * CO (wright%) Converting the O units...
  • Page 65: Appendix Ii *Software Structure Of C-Dgm

    , &&D E^ 'D 'D Appendix II *Software structure of c-DGM +---------------------+ |** PRODUCT CHOICE |CURRENT: PRODUCT1 |=====================| |PRODUCT 1 +---------------------+ |PRODUCT 2 |******* UNITS *******| |PRODUCT 3 |=====================| |PRODUCT 4 |CO2 : g/l |PRODUCT 5 Enter 3 seconds |PRESS: bar +---------------------+ +---------------------+ +--------------------------+...
  • Page 66 , &&D E^ 'D 'D Page 66 of 78 c-DGM/i-DGM...
  • Page 67: Appendix Iii Software Structure Of I-Dgm

    , &&D E^ 'D 'D Appendix III Software structure of i-DGM +---------------------+ |******* UNITS *******| |=====================| |CO2 : g/l Enter 3 seconds |PRESS: bar |TEMP: °C +---------------------+ +---------------------+ |** PRODUCT CHOICE |CURRENT: PRODUCT1 |MAIN MENU H A F F M A N S |=====================| +---------------------+ |PRODUCT 1...
  • Page 68 , &&D E^ 'D 'D Page 68 of 78 c-DGM/i-DGM...
  • Page 69: Appendix Iv Client Service

    Haffmans B.V. or your local representative or agent. Haffmans B.V. Marinus Dammeweg 30 5928 PW Venlo Holland P.O. Box 3150 NL-5902 RD Venlo Holland Telephone : (+31) 77-3232300 : (+31) 77-3232323 E-mail : NLVENService@pentair.com Internet : www.haffmans.nl c-DGM/i-DGM Page 69 of 78...
  • Page 70 , &&D E^ 'D 'D Page 70 of 78 c-DGM/i-DGM...
  • Page 71: Appendix V Warranty Conditions

    , &&D E^ 'D 'D Appendix V Warranty conditions The Haffmans B.V. warranty is supplemental to any rights the purchaser may have on the basis of the sales contract between the purchaser and the seller and on the basis of applicable laws. Haffmans B.V.
  • Page 72 , &&D E^ 'D 'D Page 72 of 78 c-DGM/i-DGM...
  • Page 73: Appendix Vi Ec Declaration Of Conformity

    , &&D E^ 'D 'D Appendix VI EC Declaration of conformity TRANSLATION OF THE ORIGINAL EC DECLARATION OF CONFORMITY EC DECLARATION OF CONFORMITY HAFFMANS B.V. Marinus Dammeweg 30 5928 PW Venlo, Holland declare entirely under our own responsibility, that the product - MEASURING, TYPE C-DGM upon this statement has reference to, is in conformity with the norms EN61326, EN55022...
  • Page 74 , &&D E^ 'D 'D Page 74 of 78 c-DGM/i-DGM...
  • Page 75: Appendix Vii Replacement Parts

    , &&D E^ 'D 'D Appendix VII Replacement parts When you order replacement parts, we request that you provide the following information with your order: Serial number of the DGM. • Position number (from the parts list below – please compare with the drawings). •...
  • Page 76 , &&D E^ 'D 'D Grease for O-rings 000.839 Accessories Power supply c-DGM & i-DGM IP67 with mains plug type EURO 101.033 Power supply c-DGM & i-DGM IP67 with mains plug type GB 101.034 Power supply c-DGM & i-DGM IP67 with mains plug type USA 101.035 Power supply c-DGM &...
  • Page 77: Appendix Viii Drawings

    , &&D E^ 'D 'D Appendix VIII Drawings Description Diagram number c-DGM overview DGM-06-3001 Key for drawing: c-DGM/01: Pos. Description Art. no. Electronics housing Measurement chamber Screw M6x8 for Calibration connection O-ring for screw M6x8 100,048 Temperature sensor Protective housing for plunger block Plunger Support bracket Communication/charging connector...
  • Page 78 , &&D E^ 'D 'D O-ring for plunger block 104,074 O-ring for measurement chamber 101,053 Plunger block c-DGM (excl. oxygen sensor) Plunger block i-DGM Screw M5x30 O-ring, plunger block, large 100,028 O-ring, plunger block, small 100,029 Page 78 of 78 c-DGM/i-DGM...