Fluoroscopy Operating Modes; Selecting The Fluoroscopy Operating Mode; Changing The Selection Of The Fluoroscopy Operating Mode - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

AXIOM Iconos R200
Examination
Fluoroscopy

Fluoroscopy operating modes

❏ Standard fluoroscopy (continuous fluoroscopy)
– In continuous fluoroscopy 30 (25)
television central control unit.
❏ Fluoroscopy with pulsed television scanning (Supervision option)
– In SUPERVISION only 15 (12.5)
case fluoroscopy is applied at half the radiation dose with half the temporal
resolution.
or
❏ Primary pulsed fluoroscopy (CAREVISION option)
– In pulsed fluoroscopy, the radiation is pulsed at 15 (12.5)
1
(3)
pulses per second. In this way a corresponding dose saving with a
lower temporal resolution is possible.
❏ Roadmap (only with DSA option) with standard fluoroscopy, with Supervision
(optional) or pulsed fluoroscopy (optional)
– Roadmap is a special fluoroscopic mode with subtraction display

Selecting the fluoroscopy operating mode

All parameters necessary for fluoroscopy are stored in organ programs.
◆ Select a corresponding organ program on the integrated generator control
console.
Changing the selection of the fluoroscopy
operating mode
It is possible to change the selection of the operating modes in connection with
the Supervision or CAREVISION options.
The current operating mode is displayed in the operating menu of the live moni-
tor.
Use the FLUOROSPOT Compact infrared control for operation.
If your system is equipped only with standard fluoroscopy, the Fluoro+ and Fluo-
ro- keys have no function.
1
Values in brackets apply to VIDEOMED DHC
AXD3-340.620.13.01.02
1
frames per second are generated by the
1
frames per second are generated. In this
1
1
or 7.5 (8.3)
or 3
7
/ 52

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents