0° Position Of The Table; Tilting The Table Up/Down - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

System remote
control console
Tableside
control panel
AXIOM Iconos R200
0° position of the table
You switch the automatic stop position of the table in the 0° position on or off
with this key.
– LED lights up: the table moves through the 0° position when tilting up/
down.
– LED does not light up: the table stops in the 0° position when tilting up/
down
With the "Stop in 0° position" function switched on, the movement stops even
if the tilting down/tilting up operating elements are operated in the 0° position.
You can continue the movement beyond the 0° position by letting go the operat-
ing element and activating it again.

Tilting the table up/down

◆ Deflect the joystick to the left/right.
– The tabletop tilts down/up.
◆ Press one of the keys.
– The tabletop tilts down/up.
AXD3-340.620.13.01.02
System Operation
System Settings
19