TriScroll 320 Series Vacuum Pump Instructions for Use General Information Storage When transporting and storing the pump, the following This equipment is designed for use by professionals. The environmental requirements should not be exceeded: user should read this instruction manual and any other...
TriScroll 320 Series Vacuum Pump Installation Requirements Shutdown Procedure 1. Close the vacuum system isolation valve. This Do not install or use the pump in an environment prevents debris in pump from being transported into exposed to atmospheric agents (rain, snow, ice), dust, the vacuum system.
TriScroll 320 Series Vacuum Pump Installation CAUTION Although the pump can pump trace particulates normally found in the Safety atmosphere, it is not designed for pro- cess solids, chemicals, powders, sol- Do not remove or modify any safety or insulating vents, condensates, or other equipment from the pump.
TriScroll 320 Series Vacuum Pump Electrical Connections Wire the motor and electrical interlocks (if applicable) in accordance with local electrical codes and the relevant electrical component manufacturer’s instructions. Table 2 lists the full load motor currents at various voltages. Table 2 Full Load Motor Currents...
TriScroll 320 Series Vacuum Pump Mechanical Connections Pump Exhaust A female 1/4" National Pipe Thread exhaust fitting is Pump Location located underneath the scroll module. This fitting swivels 360 degrees. Additionally, an NW16 male adapter with Locate the pump on a firm, level surface.
TriScroll 320 Series Vacuum Pump Overview Wiring The status of the VPI Valve is controlled by the TriScroll Using the data in Table 3, verify that the VPI Valve chosen motor internal thermal switch and its electrical supply. is compatible with the TriScroll supply voltage and The VPI Valve opens when the pump is running and frequency.
TriScroll 320 Series Vacuum Pump Troubleshooting Table 4 contains a list of possible problems, their probable causes, and corrective actions. Table 4 Troubleshooting Chart Problem Probable Cause Corrective Action Pump won’t start Circuit breaker open Close breaker. Identify cause of overload.
TriScroll 320 Series Vacuum Pump Maintenance General Information Agilent TriScroll 300 series pumps are designed to provide years of trouble-free service if maintenance procedures and intervals are observed. Bearing grease replenishment and tip seal replacement is recommended when pump base pressure has risen to an unacceptably high level for your application.
TriScroll 320 Series Vacuum Pump Factory Service Options Table 7 lists the factory-rebuild service and advance exchange of complete TriScroll Pumps, as well as factory service options that Agilent offers. Table 7 Factory Service Options Factory Service Options Part Number...
Page 17
Please follow these instructions whenever one of our products needs to be returned. 1) Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to identify all products that have pumped or been exposed to any toxic or hazardous materials.
Page 18
This page intentionally left blank. 4) HEALTH and SAFETY CERTIFICATION AGILENT TECHNOLOGIES CANNOT ACCEPT ANY PRODUCTS CONTAMINATED WITH BIOLOGICAL OR EXPLOSIVE HAZARDS, RADIOACTIVE MATERIAL, OR MERCURY AT ITS FACILITY. Call Agilent Technologies to discuss alternatives if this requirement presents a problem.
Page 19
Shipping Regulations (IATA, DOT, etc.) and carrier requirements. Customers receiving an Advance Exchange product agree to return the defective, rebuildable part to Agilent Technologies within 15 business days. Failure to do so, or returning a non-rebuildable part (crashed), will result in an invoice for the non-returned/non-rebuildable part.
Page 20
TriScroll 320 Series Vacuum Pump This page intentionally left blank.
Page 22
Service & Support North America India Southeast Asia Agilent Technologies Agilent Technologies India Pvt. Ltd. Agilent Technologies Sales Sdn Bhd Vacuum Products Division Vacuum Products Division Vacuum Products Division 121 Hartwell Avenue G01. Prime corporate Park, Unit 201, Level 2 uptown 2,...