Agilent Technologies 1168/9A User Manual page 130

Differential and single-ended probes
Table of Contents

Advertisement

4
Calibrating Probes
Calibration for Solder-In and Socketed Probe Heads
130
16
Once the skew calibration is completed, close all dialog
boxes.
17
Select the Start Skew Calibration... button and follow the
on- screen instructions.
18
Set the vertical scale for the displayed channels to 100
mV/div.
19
Set the horizontal range to 1.00 ns/div.
20
Set the horizontal position to approximately 3 ns.
21
Change the vertical position knobs of both channels until the
waveforms overlap each other.
22
Select the Setup menu choose Acquisition... from the
pull- down menu.
23
In the Acquisition Setup dialog box enable averaging. When
you close the dialog box, you should see waveforms similar
to that in
Figure
Figure 90
Overlapping Waveforms
90.
1168/9A Probes User's Guide