Process pressure measurement pressure transmitter with ceramic and metal sensors overload-resistant and function-monitored with analog, hart or profibus pa electronics (84 pages)
Solitrend MMP20 (Option D) Table of contents Table of contents Potential problems in the laboratory and at About this document ....4 the concrete plant ..... 30 Document function .
About this document Solitrend MMP20 (Option D) About this document Document function These Operating Instructions provide all of the information that is required in various phases of the life cycle of the device including: • Product identification • Incoming acceptance •...
Page 5
Solitrend MMP20 (Option D) About this document Result of a step 1, 2, 3, ... Item numbers A, B, C, ... Views Endress+Hauser...
For an overview of the scope of the associated Technical Documentation, refer to the following: • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2D matrix code (QR code) on the nameplate 1.4.1 Technical Information (TI)
Solitrend MMP20 (Option D) Basic safety instructions Basic safety instructions Requirements for personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
Basic safety instructions Solitrend MMP20 (Option D) Hazardous area To eliminate a danger for persons or for the facility when the device is used in the hazardous area (e.g. explosion protection, pressure vessel safety): ‣ Based on the nameplate, check whether the ordered device is permitted for the intended use in the hazardous area.
Solitrend MMP20 (Option D) Product description Product description The device is used to determine material moisture based on time-domain reflectometry (TDR) technology. The measuring system is suitable for mobile use (battery operation) and consists of a handheld device and a connected probe.
‣ Enter the serial number from the nameplate into the Endress+Hauser Operations App or use the Endress+Hauser Operations App to scan the 2-D matrix code (QR Code) provided on the nameplate All the information about the measuring device and the scope of the associated Technical Documentation are displayed.
Solitrend MMP20 (Option D) Electrical connection Electrical connection Connecting the probe The probe to be used is connected to the handheld device by means of a 7-pin plug. Incorrect use Only probes specially designed for this device may be connected to it. If a probe that is not designed for the device is connected to it, this may damage the device and/or the connected probe.
Operating options Solitrend MMP20 (Option D) Operating options Operating elements A0040883 2 Operating elements "Down" navigation key "Up" navigation key Display Enter key Folder key Description of the function of the keys 6.2.1 Enter key • Switch on/off: press and hold for 1 s •...
Solitrend MMP20 (Option D) Operating options 6.2.3 "Up" navigation key Previous menu item or setting: press briefly 6.2.4 "Down" navigation key • Next menu item or setting: press briefly • Clear the value memory (Average mode): press briefly Description of the icons on the display ...
Operating options Solitrend MMP20 (Option D) 10 Warning: water content values below 100 l/m are not taken into consideration or the validity of the measured value is called into question if the values vary too much. Meaning of displayed text...
Solitrend MMP20 (Option D) Commissioning Commissioning Checking the packaging contents • Handheld device • Power adapter (12 V/2 A) • Charging adapter • Protection cap • Manual • SWZ probe Charging the battery Charging the battery before using the probe for the first time 1.
Commissioning Solitrend MMP20 (Option D) Configuration and measurement For the device to be able to display the water content as a kiln-dried value with a required accuracy of ±1 to 3 l/m , the system must be set beforehand to a specific "concrete formulation characteristic"...
Solitrend MMP20 (Option D) Commissioning Special (gap-graded U) The probe measures slightly too much water and therefore has to adjust the water content downwards slightly • Very little or no gravel of size 2/8mm or 4/8mm • Standard admixtures, standard additives as well as PCEs...
Commissioning Solitrend MMP20 (Option D) Setting or changing the three concrete formulation parameters 7.7.1 Changing or entering the raw density Values must first be entered for the configurable parameters before the handheld device can switch to the measurement mode to measure the water content.
Page 19
Solitrend MMP20 (Option D) Commissioning Entering the CHAR parameter 1. Using the navigation keys , select one of the four possible options for CHAR (fine C, medium B, coarse A, or gap U) 2. Press Enter to confirm your entry 7.7.3...
Commissioning Solitrend MMP20 (Option D) 12 The Meas menu Set new parameters Pressed briefly: delete last individual measurement value; pressed for longer: delete entire measurement series Conductivity / assessment of the cement Standard deviation: more single measurements are needed if the std-dev is > 0.5 !
Solitrend MMP20 (Option D) Commissioning safety and reliability when monitoring and checking a known type of concrete. It is advisable for the user to document the particular varieties of concrete that are measured. This makes it easier to verify the values during subsequent control measurements.
Page 22
Commissioning Solitrend MMP20 (Option D) The serial number of the probe appears on the display as soon as the connection has been established successfully. "Probe not found" appears on the display if a connection cannot be established. No connection established with the probe despite multiple attempts ‣...
Page 23
Solitrend MMP20 (Option D) Commissioning 1. Select the desired automatic switch-off time by pressing the navigation key several times 2. Select the desired brightness of the display or switch it off completely by pressing the navigation key several times 3. Press Enter to activate and save the selected settings ...
Page 24
Commissioning Solitrend MMP20 (Option D) The sensitivity of the concrete measurement can be changed by selecting another calibration curve. The standard calibration curve "Cal. No.: 4" is set by default for concrete. • Do not change this setting, or only change this setting if you are measuring a material other than fresh concrete •...
Solitrend MMP20 (Option D) SWZ probe SWZ probe Introduction The SWZ probe uses radar technology at 1 GHz and a probe whose measuring field penetrates deep into the material to be measured. Plastic and liquid fresh concrete with a consistence class of F2 to F6 can be measured by hand easily and directly. An automatic averaging function, with 4 to 10 single measurements taken, ensures a representative measurement of the material mix.
SWZ probe Solitrend MMP20 (Option D) A0040932 16 Using the SWZ probe Correct use of the probe: • The probe measuring field must be fully located in the concrete • The probe head must be fully inserted in the concrete to be measured, without any "air gaps"...
Page 27
Solitrend MMP20 (Option D) SWZ probe Please note the following: • There should not be any residue of old concrete around the ceramic on the probe surface. Clean the surface with a wire brush if necessary. • The amount of concrete in the bucket should be at least 3 cm higher than the length of the probe head (<18 cm).
Page 28
SWZ probe Solitrend MMP20 (Option D) 8. Repeat the process 4 to 5 times in total A0040937 In the case of concrete that "sticks", the dark ceramic surface of the probe should be wiped clean before each measurement to ensure that concrete residue "sticking" to the surface of the probe does not falsify the measurement.
Page 29
Solitrend MMP20 (Option D) SWZ probe 2. At the edge of the bucket, vertically insert the probe head with the plug-in blade (plastic) fully into the concrete. A0040931 The plug-in blade ensures that the larger gravel pieces do not "drift away" from the probe head towards the side during the measurement, which can cause inaccuracies.
SWZ probe Solitrend MMP20 (Option D) Potential problems in the laboratory and at the concrete plant 8.4.1 Situation 1: Mixing the concrete with dry aggregates Depending on the rock, it can take some time for dry aggregates to become saturated after the mixing process.
Page 31
Solitrend MMP20 (Option D) SWZ probe 2. The concrete sample with a normal-distribution grading curve and a water target value of 170 l/m³ was measured with the SWZ probe and 170 l/m³ was displayed as the reading. 3. After that, a 5 kg concrete sample was kiln-dried. A kiln-dried value of 149 l/m determined, i.e.
Commissioning fresh concrete measurement Solitrend MMP20 (Option D) Commissioning fresh concrete measurement NOTICE During the measurement process, there should never be any metal parts in the vicinity of the probe head, as metal can affect the measuring field of the probe. Fresh concrete should always be measured in a plastic bucket, as this rules out any influence of metal on the measurement.
Page 33
Solitrend MMP20 (Option D) Commissioning fresh concrete measurement coarse, normal or special" with the CHAR parameter (see the "Settings" chapter). The device is set to the variety of concrete with the corresponding rock type with the G-Set parameter. G-Set is entered in liters/m3 and can be entered in increments of one liter up to ± 50 l.
Page 34
Commissioning fresh concrete measurement Solitrend MMP20 (Option D) 1. Make sure that no air is trapped in the fresh concrete 2. Press the Enter key to "Start measurement". The water content, determined using the raw density, is calculated in l/m and displayed.
Solitrend MMP20 (Option D) Commissioning fresh concrete measurement 9.1.4 Starting the next individual measurement To prevent the segregation of the concrete, it is recommended to mix the fresh concrete again after 5 measurements. With regard to the representativeness, this simply involves changing the material mix or composition with different sized gravel pieces at the probe head.
Commissioning fresh concrete measurement Solitrend MMP20 (Option D) The SWZ probe also measures additives that behave like water during the measurement process. This must also be taken into consideration when evaluating and determining the w/c water content. Applicable value for the core water for the calculation when kiln-drying:...
Solitrend MMP20 (Option D) Commissioning fresh concrete measurement • When weighing the dry kiln-dried sample, pay attention to the temperature of the kiln- dried sample. When weighing a very hot kiln-dried sample, the upward lift of air streams can produce a considerable error on the scales. For example, when weighing a 4 kg weight, the hot air stream can produce a difference of 30 g, depending on the weighing unit.
Commissioning fresh concrete measurement Solitrend MMP20 (Option D) The SWZ probe measures the same parts of water as the kiln-drying method: • The free water in the concrete mix which enters into the calculation of the w/c-ratio. This water is the actual value that is sought when using the SWZ probe.
Solitrend MMP20 (Option D) Managing and archiving concrete formulations Managing and archiving concrete formulations With the correct settings for the CHAR parameter (fine, coarse, normal, special) and G-Set parameter (core water and additives), the measurement results of the SWZ probe should correlate relatively well with verified actual values or target values.
S1 moisture probe Solitrend MMP20 (Option D) S1 moisture probe The handheld device can be used in conjunction with the S1 probe to measure moisture in sand, gravel and other bulk solids. 11.1 Connecting the S1 probe 1. Connect the S1 probe to the handheld device ...
Solitrend MMP20 (Option D) S1 moisture probe 1. Press the Enter key briefly to start the measurement The device starts measuring and a rotating symbol appears in place of the battery symbol in the top right-hand corner for the duration of the measurement process.
Page 42
S1 moisture probe Solitrend MMP20 (Option D) • Auto-power-off: setting for automatic switch-off • Display lighting: setting for background lighting • Switch-off time • Brightness • Display contrast: setting for the optimum contrast • Probe info: displays information about the probe •...
Page 43
Solitrend MMP20 (Option D) S1 moisture probe 2. Press the Enter key This activates the selected calibration Afterwards, the symbol appears in the top right-hand corner of the display indicating that the option has been activated. In addition, the "!" symbol appears in front of the active calibration.
Page 44
S1 moisture probe Solitrend MMP20 (Option D) 3. Select the percentage moisture using the navigation keys 4. Press the Enter key The setting is accepted 5. Press the Enter key again The material measurement process is started Four measurements are taken in order to increase the accuracy.
Page 45
Solitrend MMP20 (Option D) S1 moisture probe higher-value polynomial is useful to achieve better accuracy, the linear equation often suffices, particularly in the lower moisture range, and delivers very good results. Two material samples with different moisture values are required to perform a 2- point material calibration.
Page 46
S1 moisture probe Solitrend MMP20 (Option D) 8. Press the Enter key again The material measurement process is started Four measurements are taken in order to increase the accuracy. An average of these measurement values is then calculated. The measuring time is approximately 20 seconds.
Solitrend MMP20 (Option D) S1 moisture probe 11.4 Using the S1 probe 11.4.1 Measuring volume In theory, the electric and magnetic field lines penetrate the material to be measured to an infinite depth. However, the effective penetration depth of the S1 probe, which is relevant for measurement, is approximately 80 mm (3.15 in) (twice the distance between the...
Page 48
S1 moisture probe Solitrend MMP20 (Option D) A0040890 The container must have a volume of 10 l or more and must be of a non-metal material >10 l <10 l A0040891 The container must be roughly cylindrical A0040892 The level of product in the container must be at least 5 cm greater than the probe rod...
Page 49
Solitrend MMP20 (Option D) S1 moisture probe Take the measurements according to the following procedure: 1. Fill the container with sand A0040894 2. Lift the container approximately 5 cm and then drop it. Repeat this five times (more if necessary).
Page 50
S1 moisture probe Solitrend MMP20 (Option D) 4. Take the measurement with the handheld device 5. Remove the probe from the sand and shake the sand to loosen it again 6. Repeat steps 2 to 4 two more times so that you have 3 measured values in total ...
Page 51
Solitrend MMP20 (Option D) S1 moisture probe In the case of the S1C probe, you can replace the rods as follows: 1. Unscrew the probe rods from the probe body A0041449 2. Remove the probe rods A0041450 3. Remove the sealing rings ...
Page 52
S1 moisture probe Solitrend MMP20 (Option D) 5. Insert the probe rods A0040880 6. Screw the probe rods into the probe body A0041452 Endress+Hauser...
Solitrend MMP20 (Option D) Technical data Technical data 12.1 Handheld device • Height: 36 mm • Width: 64 mm • Length: 150 mm • Weight: (with battery) approx. 437 g • Power demand: • Power down: 35 µA • Idle:...
Page 54
Technical data Solitrend MMP20 (Option D) 130 (5.12) 110 (4.33) 135 (5.31) 155 (6.1) 218 (8.58) A0040884 • Power supply: 12 to 24 V • Current consumption: 100 mA @ 12 V during 2 to 3 s measuring cycle time •...