Load Effect - UNI-T UTL8500 Series User Manual

Programmable dc electronic load
Table of Contents

Advertisement

2. Move the cursor to the needed setting position by pulse knob, and press the Enter
button
to change the parameter (the parameter changes from yellow to white) by
keyboard or pulse knob. Press the Enter button
set value changes from white to yellow.
3. Other parameters can be modified in a similar way.
4. Press the ON button, the electronic load starts to load, and the indicator light below the
button lights up. If you need to stop the load, press the ON button again, and the
running indicator light goes out.
5. Press [Curve] to view the curve of battery discharge when the test is over. (Note:
User can't view the curve during the test.)

4.2.7 Load Effect

The load effect mode can carry out loading under 3 different load conditions (10%, 50%,
and 100%), and continue the load for a preset time (5s), and then record the voltage under
different loads. Finally, calculate the load regulation rate Regulation, △V and other values
according to the formula.
Parameter Setting in Load Effect Mode:
Parameter
Rate Voltage
Rate Current
Formula: Vmax=Vdc@Imin, Vmin=Vdc@Imax, △V=Vmax-Vmin, Regulation=△V/Vset
Before the load effect test, input the rated voltage and rated current, and then press ON to
enter the test. At this time, the load will be loaded according to 3 different loads in turn, and
the values of △V and Reg will be detected. During the test, the ON button can be used to
control the load switch.
Figure 4-2-7-1 Interface of Load Effect Mode
UTL8500 Series User Manual
again to confirm. At this time, the
Description
Set the Rated Voltage of the Load
Set the Rated Current of the Load
33