Interpretation of results

The measured value is shown on the display. The bullet point under traffic light scale ranks the result relative to the
patient's threshold references.
INFORMATION
Interpretation of the results and setting of thresholds must be advised by your physician
or healthcare professional.
Default values set for the traffic light scale for adults in accordance with the ATS Guideline 2011,
see "Literature" in chapter 1.3:
Green: < 25 ppb, Yellow: 25 ppb - 50 ppb, Red: > 50 ppb
For children below 12 years ATS guideline defines:
Green: < 20 ppb, Yellow: 20 ppb - 35 ppb, Red: > 35 ppb
If you have further questions, contact your distributor or the Vivatmo me service center.
2.5

Displaying stored measured values

The Vivatmo me automatically stores up to 1,000 measured FeNO values identified by time, date, and ID number. The
measured values are stored in descending order starting with the latest. You can retrieve the measured values anytime.
• Press one of the ARROW buttons to show older measured values.
• Press the ENTER button to exit the view.
Instructions for use
Measurement of FeNO in exhaled breath