Product description
Model
Measuring range
Accuracy
Precision
Lifetime, device service life,
memory capacity
Power source
FeNO monitoring system
Vivatmo me, catalogue number F09G100149
5 ppb to 300 ppb
± 5 ppb below 50 ppb, ±10 % for ≥50 ppb, ± 15 % for ≥ 160 ppb
expressed as the upper/lower confidence limit of 95 %
± 5 ppb below 50 ppb, ± 10 % for ≥50 ppb, ± 15 % for ≥ 160 ppb expressed as
one standard deviation for replicate measurements with the same instrument
1,000 measurements
4 AAA batteries 1.5 V, Useful life of primary batteries:
Lithium/iron disulfide up to 60 attempts, alkaline up to 25 attempts,
(Battery and compartment lid are detachable parts)