Materials And Solvents - Agilent Technologies Varian 320-MS Pre-Installation Instructions

Quadrupole mass spectrometer lc/ms
Hide thumbs Also See for Varian 320-MS:
Table of Contents

Advertisement

Materials and Solvents

The Varian Representative will prepare solutions to tune and evaluate the LC/MS. Please
supply the materials on Table 13 and the solvents on Table 14.
Table 13 Materials
Quantity (each)
Table 14 Solvents for ESI or APCI
Quantity (each)
Additional tubing is required to complete installation. Most of this tubing is included in the
accessory kits of the LC modules. Additional tubing (PEEK™ or stainless steel) may be
required for installation of special valves or modules. For most analytical HPLC systems, use
1/16 in. tubing with an ID of 0.005 in. Use 0.005 in. ID or smaller tubing downstream of the
sample injector and autosampler to prevent peak broadening. Keep tubing as short as possible
to prevent peak broadening and to minimize run time.
9300025700: Rev. A
Item
2
50 mL clean and new volumetric flasks
2
1 L clean and new bottles for mobile phase reservoirs
1
1 L or larger bottle for LC/MS waste
2
10 mL clean and new volumetric flask
2
100 mL clean and new volumetric flask
1
Pipette 100 to 1000 µL and tips
1
Pipette 20 to 200 µL and tips
1
50 mL clean and new volumetric flask (for optional APCI)
1
500 mL clean and new volumetric flask (for optional APCI)
Solvent
2
4-liter bottle HPLC grade or better water, new and unopened
2
4-liter bottle HPLC grade or better, acetonitrile, new and unopened
Check the checklist box: Materials and solvents of specified grade are on site.
12 of 14
Pre-installation Instructions

Advertisement

Table of Contents
loading

Table of Contents