Argon; Materials And Solvents - Agilent Technologies Varian 325-MS LC/MS Pre-Installation Instructions

Quadrupole mass spectrometer
Table of Contents

Advertisement

Argon

Argon is required as a collision gas for MS/MS work. The argon supply uses 1/8 in. connectors.
Use a two-stage 0-100 psi pressure regulator with a stainless steel diaphragm. The purity of the
argon must be greater than 99.0%.

Materials and Solvents

The Varian Representative will prepare solutions to tune and evaluate the LC/MS. Please
supply the materials on Table 11 and the solvents on Table 12.
Table 11 Materials
Quantity (each)
Table 12 Solvents for vESI or APCI
Quantity (each)
9300015500 Rev. C
Check the checklist box: Argon (greater than 99.0%), regulator, and gas lines
are installed.
Material
2
50 mL clean and new volumetric flask
2
1 L clean and new bottle for mobile phase reservoirs
1
1 L or larger bottle for LC/MS waste
2
10 mL clean and new volumetric flask
2
100 mL clean and new volumetric flask
1
Pipette 100 to 1000 µL and tips
1
Pipette 20 to 200 µL and tips
1
50 mL clean and new volumetric flask (for optional APCI)
1
500 mL clean and new volumetric flask (for optional APCI)
Solvent
2
4-liter bottle LC/MS grade or better water, new and unopened
2
4-liter bottle LC/MS grade or better, acetonitrile, new and unopened
11 of 14
Pre-installation Instructions

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents