Front/Rear Panel
Front Panel
Front Panel
Front Panel Features
Figure 3-1
Port Connection
Status
Ground
Line Power
Connector
Switch
Line Power Switch
The test set line power switch is located at the bottom left corner of the
front panel. See
the test set either On (
The front panel line switch disconnects the mains circuits from the
mains supply after the EMI filters and before others parts of the
instrument.
Ports 1—8
PORTS 1 through 8 are female Type-N 75 Ω connectors that are used to
connect to the device under test.
Never attempt to use 50 Ω Type-N cables, connectors, or devices with a
CAUTION
female Type-N 75 Ω connector. Irreparable damage will occur.
Do not input more than +20 dBm or 25 Vdc to these ports or damage to
CAUTION
the internal RF switches or the analyzer will occur.
38
Reflection
Connector
Ports 1-8
Figure
3-1. The line power switch turns the power to
|
O
) or Off (
).
Transmission
Connector
Chapter 3