Front Panel; Line Power Switch; Ports 1-8 - Agilent Technologies 87075C Option H08 User's And Service Manual

Hide thumbs Also See for 87075C Option H08:
Table of Contents

Advertisement

Front/Rear Panel

Front Panel

Front Panel
Front Panel Features
Figure 3-1
Port Connection
Status
Ground
Line Power
Connector
Switch

Line Power Switch

The test set line power switch is located at the bottom left corner of the
front panel. See
the test set either On (
The front panel line switch disconnects the mains circuits from the
mains supply after the EMI filters and before others parts of the
instrument.
Ports 1—8
PORTS 1 through 8 are female Type-N 75 Ω connectors that are used to
connect to the device under test.
Never attempt to use 50 Ω Type-N cables, connectors, or devices with a
CAUTION
female Type-N 75 Ω connector. Irreparable damage will occur.
Do not input more than +20 dBm or 25 Vdc to these ports or damage to
CAUTION
the internal RF switches or the analyzer will occur.
38
Reflection
Connector

Ports 1-8

Figure
3-1. The line power switch turns the power to
|
O
) or Off (
).
Transmission
Connector
Chapter 3