Technical Specifications

Performance
Maximum heater plate temperature: 167°F
Temperature cut-out: 199°F
Maximum output humidity: 95% relative humidity (control dial set to maximum
setting, flow rate 50 L/min, ambient temperature of 68°F, ambient relative
humidity of 40%).
Dimensions (H x W x D)
4.8 in x 7.2 in x 8.6 in (docking station and water chamber)
Weight
2 lb (docking station and unfilled water chamber)
Water Capacity
To maximum fill line: 400 mL
Power Supply
Input range for AutoSet Spirit Unit with HumidAire 2i: 100–240 V, 50–60 Hz,
<528 VA (maximum power consumption)
Actual power consumption will vary, depending on factors such as temperature
setting, set pressure, the use of accessories, height above sea level and ambient
temperature.
Housing Construction
Docking station: Flame retardant engineering thermoplastic
Water chamber: Injection molded plastic
Environmental conditions
Operating temperature: +41°F to +104°F
Operating humidity: 10–95% non-condensing
Storage and transport temperature: -4°F to +140°F
Storage and transport humidity: 10–95% non-condensing
IEC 60601-1 Classification
Class II (double insulation)
Type CF
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments.