  • Page 2: Indications For Use

    Indications for use AirSense 10 AutoSet for Her The AirSense 10 AutoSet for Her self-adjusting device is indicated for the treatment of obstructive sleep apnoea (OSA) in female patients weighing more than 30 kg. It is intended for home and hospital use.
  • Page 3: Device Overview

      At a glance The AirSense 10 includes the following:  Device  Air tubing  Power supply unit  Travel bag  SD card (already inserted). Contact your care provider for a range of accessories available for use with the device including: ...
  • Page 4: About The Control Panel

    3. Connect the air tubing firmly to the air outlet located on the rear of the device. 4. Connect the free end of the air tubing firmly onto the assembled mask. See your mask user guide for detailed information. Recommended masks are available on www.resmed.com. English...
  • Page 5: Power Save Mode

    Power save mode Your AirSense 10 device records your therapy data. In order to allow it to transmit the data to your care provider, you should not unplug the device. However, you can put it into power save mode to save electricity.
  • Page 6: My Options

    The Home screen is displayed. My Options Your AirSense 10 device has been set up for your needs by your care provider, but you may find you want to make small adjustments to make your therapy more comfortable. Highlight My Options and press the dial to see your current settings.
  • Page 7: More Options

    Caring for your device It is important that you regularly clean your AirSense 10 device to make sure you receive optimal therapy. The following sections will help you with disassembling, cleaning, checking and reassembling your device.
  • Page 8: To Replace Air Filter

    2. Connect the free end of the air tubing firmly onto the assembled mask. Therapy data Your AirSense 10 device records your therapy data for you and your care provider so they can view and make changes to your therapy if required. The data is recorded and then transferred to your care provider wirelessly or via an SD card.
  • Page 9: To Remove Sd Card

    Note: The SD card should not be used for any other purpose. Travelling You can take your AirSense 10 device with you wherever you go. Just keep the following in mind:  Use the travel bag provided to prevent damage to the device.
  • Page 10: General Troubleshooting

    Troubleshooting If you have any problems, have a look at the following troubleshooting topics. If you are not able to fix the problem, contact your care provider or ResMed. Do not try to open the device. General troubleshooting Problem/possible cause...
  • Page 11: Device Messages

      Device messages Device message/possible cause Solution High leak detected, connect your tubing Air tubing may not be connected properly. Make sure the air tubing is firmly connected at both ends. Mask may be fitted incorrectly. Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
  • Page 12: General Warnings And Cautions

     Do not perform any maintenance tasks while the device is in operation. CAUTION  Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the effectiveness of the treatment and/or damage the device.  Use only vented masks recommended by ResMed or by the prescribing doctor with this device.
  • Page 13: Technical Specifications

    5 to 95% relative humidity, non-condensing Electromagnetic compatibility The AirSense 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1- 2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device.
  • Page 14 Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust Aircraft use ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel. Wireless module Technology used: 2G GSM It is recommended that the device is a minimum distance of 2 cm from the body during operation.
  • Page 15   Notes:  The manufacturer reserves the right to change these specifications without notice.  The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should not be fitted to the mask. ...
  • Page 16: Limited Warranty

    ResMed. ResMed recommends that the AirSense 10 device be inspected and serviced by an authorised ResMed Service Centre if there is any sign of wear or concern with device function. Otherwise, service and inspection of the products generally should not be required during their design life.
  • Page 17: Further Information

     Portable diagnostic/screening devices This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
  • Page 18 See www.resmed.com for other ResMed locations worldwide. For patent information, see www.resmed.com/ip. Air10, AirSense, AutoSet, ClimateLine, HumidAir and SlimLine are trademarks of ResMed Ltd. AutoSet, SlimLine and ClimateLine are registered in U.S. Patent and Trademark Office. SD Logo is a trademark of SD-3C, LLC. © 2014 ResMed Ltd. 378129/1 2014-07 ResMed.com...