Problem/possible cause
Displays error message: Heated tube fault, replace tube
Device has been left in a hot environment.
There is a fault in your ClimateLine air tubing.
Displays error message: Humidifier fault, replace humidifier
There is a fault in your H5i heated humidifier.
Device has been left in a hot environment.
Displays error message: Tube blocked, please check your tube
Air tubing is blocked.
Water splashing on your face from the H5i
The water tub is overfilled.
Condensation is forming in the air tube and mask.
Leaking water tub
The water tub may be damaged or cracked.
The cleanable water tub is not assembled
correctly.
Air feels too warm/cold in the mask
The temperature of the ClimateLine air tubing is
set too high/low.

Technical specifications

Maximum heater plate temperature
Temperature cut-out
Maximum gas temperature
Nominal dimensions (L x W x H)
Weight (Water tub)
Weight (Cleanable water tub)
Water capacity
Docking station
Cleanable water tub
Water tub
Operating temperature
Operating humidity
Storage and transport temperature
2
Solution
Allow to cool before re-use. Disconnect the power cord and then
reconnect it to restart the device.
Discontinue using your ClimateLine air tubing and contact your clinician/
service provider. Use Standard or SlimLine tubing in the interim.
Discontinue using your H5i and contact your clinician/service provider.
Allow to cool before re-use. Disconnect the power cord and then
reconnect it to restart the device.
Check your air tubing and remove any blockages.
Check that the water level is below the maximum water level mark.
Turn the H5i or relative humidity setting down via the menu.
Contact your service provider for a replacement.
Check for damage and reassemble the cleanable water tub correctly.
Turn up/down the heated tubing temperature via the menu.
65°C
74°C
≤ 41°C
Docking station and water tub: 153 mm x 145 mm x 86 mm
Docking station and unfilled water tub 0.67 kg
Docking station and unfilled water tub 0.77 kg
To maximum fill line 380 mL
Flame retardant engineering thermoplastic, aluminium
Injection molded plastic, stainless steel and silicone seal
Injection molded plastic, aluminium and thermoplastic elastomer
º
º
+5
C to +35
C
10–95% non-condensing
º
º
-20
C to +60
C