BALANCE NETWORK
300N
0.1 %
300n
0.1 %
1501
Set the Wave Analyzer controls as follows :
MAX . INPUT VOLTAGE
REF .
ADJ .
Switch
METER .
RANGE
MODE
BANDWIDTH
FREQUENCY
FREQUENCY RANGE
SWEEP
SWEEP RATE
C.
Adjust the Wave Analyzer COARSE for maximum meter indication , and
adjust the REF ADJ for a 0dB meter indication .
Reset Wave Analyzer controls as follows :
Set the Wave Analyzer SWEEP to START and observe the meter indication
as the Wave Analyzer sweeps to 400 Hz ,
Reset the SWEEP RATE and BANDWIDTH at the frequencies shown in
Table 5-5 .
The meter indication should be down greater than 80 dB across the full
sweep range ,
except for the 100 kHz fundamental and its harmonics.