Download Print this page

HP 204C Technical Manual page 28

Hide thumbs Also See for 204C:

Advertisement

LITIID )
Convert the reading on the DC NULL VOLTMETER to the percentages
listed in Table 5-3 by the following procedure .
Divide the DC NULL
VOLTMETER indication by the THERMAL CONVERTER output voltage
recorded in step c of this paragraph .
Multiply this value by 100 to get
percent of output change .
Divide this percentage by the Calibration Report
multiplier to obtain a percentage within the tolerances listed in Table 5-3 .
Example :
100 kHz
Frequency
THERMAL CONVERTER output
DC NULL VOLTMETER reading
Calibration Report multiplier
Table 5-3 tolerance
0.025
5 5
=
0.005
0.005 x 100
=

Advertisement