Equipment Requirements

The Tester's recommended calibration cycle is one year. Table 2 lists the calibration
equipment required for the Tester. If the recommended models are unavailable, you can
substitute equipment with equivalent specifications.
For safe operation and maintenance of the product, have an
approved technician repair the product.
Equipment
Multifunction Electrical Tester
Calibrator
Decade Resistor Box
PC
IR adapter Cable
IR Cable adapter holder
Zero Adapter
Warning
Table 2. Equipment Requirements
Description
AC Voltage Range: 0 to 500 V ac
Accuracy: ±0.215 %
Frequency Range: 50 Hz
Insulation Resistance: 10 kΩ to 900 MΩ
Accuracy:
10 kΩ to 200 MΩ = ±0.413 %
200 MΩ to 900 MΩ = 2.5 %
Low Resistance: 2 Ω to 1800 Ω
Accuracy: ±0.416 %
Loop Resistance: 100 mΩ to 1.8 kΩ
Accuracy:
100 mΩ = ±5 %
1.8 Ω = ±1 %
18 Ω to 180 Ω = ±0.5 %
500 Ω to 1800 Ω = ±1.5 %
RCD
Trip Current Measurement
Range: 10 to 1000 mA
Accuracy:
0.5 x 1 and 1 x 1 mode = ±2.5 %
1.4 x 1 and 2 x 1 mode = ±2.5 %
5 x 1 mode = ±2.5%
Trip time Measurement
Range: 30.0 to 500.0 ms
Accuracy: ±1.2 %
Resistance: 18 Ω, 180 Ω, 1800 Ω
Accuracy: 0.75 % or better
Power: 0.5 W minimum
PC with Windows 98 or newer
---
FLK-1653-2014-OSP, 2043365, IR Adapter
Fluke-165XB-8001, Zero Adapter
Electrical Installation Tester
Equipment Requirements
Recommended
Model
Fluke 5320A Calibrator
GenRad 1433
---
Fluke P/N 1590638
Fluke P/N 3270172
Fluke P/N 3301338