Operating Ranges and Uncertainties per EN 61557

Function
Display Range
V
0.0 V ac – 500 V ac
EN 61557-1
R LO
0.00 Ω - 2000 Ω
EN 61557-4
R ISO
0.00 MΩ - 1000 MΩ
EN 61557-2
Z
(No Trip)
I
0.00 Ω - 2000 Ω
Z
(Hi Current)
I
Z I
0.00 Ω - 2000 Ω
EN 61557-3
Z
(Hi Current, Hi Res)
I
0 mΩ - 9999 mΩ
R
E
0.00 Ω - 2000 Ω
∆ T
0.0 ms – 2000 ms
∆ T, I ∆N
I
∆N
EN 61557-6
3 mA – 550 mA
(VAR 3 mA – 700 mA)
R E
0.0 Ω - 2000 Ω
EN 61557-5
Phase
EN 61557-7
EN 61557
Measurement
Range Operating
Uncertainty
50 V ac – 500 V ac
U
= 230/400 V ac
N
±(2% + 2 dgt)
f = 50/60 Hz
4.0 V dc <U
0.2 Ω - 2000 Ω
±(10 % + 2 dgt)
R
≤2.00 Ω I
LO
1 MΩ - 200 MΩ
U
= 50 / 100 / 250 / 500 /
N
±(10 % + 2 dgt)
1000 V dc
200 MΩ - 1000 MΩ
I
= 1.0 mA
N
±(15 % + 2 dgt)
0.4 Ω - 2000 Ω
±(15 % + 6 dgt)
0.2 Ω - 200 Ω
U
= 230/400 V ac
N
±(10 % + 4 dgt)
f = 50/60 Hz
100 mΩ - 9999 mΩ
I
= 0 A – 10.0 kA
K
±(8 % + 20 dgt)
10 Ω - 1000 Ω
±(10 % + 2 dgt)
25 ms – 2000 ms
∆ T = 10 / 30 / 100 / 300 / 500 /
±(10 % + 1 dgt)
1000 / VAR mA
3 mA – 550 mA
I
= 10 / 30 / 100 / 300 /
∆N
±(10 % + 1 dgt)
500 / VAR mA
10 Ω - 2000 Ω
f = 128 Hz
±(10 % + 2 dgt)
1 : 2 : 3
Nominal Values
<24 V dc
Q
≥200 mA
N
Electrical Installation Tester