Performing A Biocel Ready-State Check - Agilent Technologies BioCel 1200 User Manual

Table of Contents

Advertisement

Related topics

Performing a BioCel ready-state check

Ready state
conditions
Related topics
4. Make sure that all four lights on the power panel are on.
If you have not previously selected the option to sense the air
pressure automatically, check that the operating pressure is 100–
110 psi (0.62–0.69 MPa) for the main regulator and 35–60 psi (0.27–
0.34 MPa) for the robot regulator.
5. Turn on the devices if they are not already on.
For more information, see the device's user guide.
For information about...
Connections for utilities
Power panel
Vacuum system
BioCel hazard warnings
When the BioCel is on and ready for a protocol to be opened:
The status lights are off. (BioCel 1200 does not have status lights.)
All devices are on, pneumatic devices are supplied with air, and a
vacuum is supplied where used.
The computer is on and ready for a user to log in to the operating
system.
The UPSs are charged, ready to provide battery backup if the main
AC power supply is cut.
All doors are closed.
The interlock override is off so that opening an enclosure door or
pressing an emergency button will activate the emergency stop
circuit.
For more information about...
Status lights
Turning on the BioCel
Chapter 3: Preparing for a run
See...
"Utilities connections" on page 19
"Power panel" on page 21
"Vacuum system" on page 31
"Safety information about the BioCel"
on page 8
See...
"Monitoring the status of a run" on
page 63
"Turning on the BioCel" on page 43
45
BioCel User Guide