Download Print this page

Fluke 8808A Manual Supplement page 2

Hide thumbs Also See for 8808A:

Advertisement

8808A Getting Started
Change #1,
44931, 46479, 47892
On page 1, in the Specifications section, under Relative Humidity (non-condensing), replace
Operating with the following:
Operating ................................................ <90 % (0 C to 28 C)
On page 5, under DC Current, Shunt Resistance, replace the Burden Voltage with the
following:
Burden voltage <5 mV for 200 A and 2 mA range.
In the Input Characteristics table, change the Burden Voltage for both 200 µA and 2 mA
ranges,
From:
<1 mV
To:
<5 mV
Change #2
On page 6, under the Warning, replace steps 4 and 5, with the following:
4. Insert the other end of the test lead into the mA terminal. If the fuse is good, the Meter
displays a reading of <10 . If the fuse is blown, the Meter displays  to indicate an overload.
5. Remove the test lead from the mA terminal and insert it into the 10 A terminal. If the fuse is
good, the Meter displays a reading of <2 . If the fuse is blown, the Meter displays  to
indicate an overload.
Change #3,
53744
Under the LIMITED WARRANTY AND LIMITATION OF LIABILITY replace the second
sentence with the following:
The warranty period is three years limited and begins on the date of shipment.
On page 11, replace Table 6 with the following:
Premium Test Lead Set
Fuse, .25*1.25, 0.063 A, 250 V, Slow
Fuse, .25*1.25, 0.125 A, 250 V, Slow
F1 - Fuse, 11 A, 1000 V, Fast, 406INX1.5IN, BULK
F2 - Fuse, 440 mA, 1000 V, Fast, 406INX1.375IN, BULK
Rack Mount Kit 8845A & 8846A Single
Rack Mount Kit 8845A & 8846A Dual
USB to RS-232 Cable Adapter (included)
Precision Electronic Probe Set
2X4 Wire Ohms 1000 V Test Lead
FlukeView Forms Basic Software (included)
FlukeView Forms Software Upgrade to enhanced version
2/10
<75 % (28 C to 40 C)
<45 % (40 C to 50 C)
Table 6. Accessories
Item
1
Manual Supplement
Model / Part Number
TL71
163030
166488
803293
943121
Y8846S
Y8846D
884x-USB
TL910
TL2X4W-PTII
FVF-SC5
FVF-UG