Section 2: Mechanical Diagnostics

Transport Malfunction

Possible Cause
RACK and CROSSOVER
ASSEMBLIES
ROLLER ASSEMBLIES
DRYER
TOP COVER
5B6342 – January 1996
RACK and CROSSOVER ASSEMBLIES:
correct positions
seated correctly
• squareness
– See the Adjustments and Replacements, Publication No. 5B6340,
RACKS and CROSSOVERS.
Adjusting the Squareness of the CROSSOVERS
Adjusting the Squareness of the RACKS
• cleaned completely
– See the Preventive Maintenance, Publication No. 5B6747.
CROSSOVER TROUGHS
• correct positions
ROLLERS
• correct positions
• rotate freely
• GUDGEONS
– no damage
– If necessary, install new ROLLERS.
GEARS, SPROCKETS, and IDLERS
• engage correctly
• not broken or worn
BEARINGS
• no wear
SPRINGS and E-RINGS
• not broken or missing
RACK ASSEMBLY
• DRIVE CHAIN
– tension
AIR TUBE BAFFLES
• installed
Temperature setting of the DRYER
lowest possible setting to provide the best image quality
DRYER ROLLERS and AIR TUBES
• seated correctly
DRYER BELT
• moves correctly
TOP COVER
• closed