4.
End the session using the niScope Close VI.
LabVIEW VI
Verification
After the 15-minute warm-up time, always self-calibrate the NI 5114 before
Note
beginning a verification procedure.
This section describes the program you must write to verify either the calibration test limits or
the published specifications for the NI 5114.
If any of these tests fail after you perform an external adjustment, make sure
Note
that you have met the requirements listed in the
Procedures
Vertical Offset and Vertical Gain Accuracy
Table 4 contains the input parameters for verifying both vertical offset accuracy and vertical gain
accuracy of the NI 5114.
To verify vertical offset accuracy, complete the procedures described in the
Accuracy
section for each of the 20 iterations listed in Table 4 for channel 0, then repeat the
procedures for channel 1. The Calibration Test Limits and Published Specifications for vertical
offset accuracy are shown in Table 5.
To verify vertical gain accuracy, complete the procedures described in the
Accuracy
section for each of the 20 iterations listed in Table 7 for channel 0; then repeat the
procedures for channel 1. The Calibration Test Limits and Published Specifications for vertical
gain accuracy are shown in Table 6.
Table 4. NI 5114 Input Parameters for Vertical Offset Accuracy
Iteration
1
2
3
4
8 | ni.com | NI 5114 Calibration Procedure
sections before you return the device to National Instruments for repair.
and Vertical Gain Accuracy Verification
Max Input Frequency (MHz)
C/C++ Function Call
Call
niScope_close
following parameter:
vi: The instrument handle from
niScope_init
Test Equipment
125
125
125
125
with the
and
Calibration
Vertical Offset
Vertical Gain
Range
0.04
0.1
0.2
0.4