12. End the session using the niScope Close VI.
LabVIEW VI
Table 8. NI 5114 Timing Accuracy Limits and Published Specifications
Calibration Test Limits
±5.3 ppm
Bandwidth and Flatness
Complete the following steps to verify the bandwidth and flatness of the NI 5114. You must
verify both channels with each iteration listed in Table 9.
Verification of bandwidth is optional.
Note
1.
Open a session and obtain a session handle using the niScope Initialize VI.
LabVIEW VI
NI 5114 Calibration Procedure | © National Instruments | 27
C/C++ Function Call
Call
niScope_close
following parameter:
vi: The instrument handle from
niScope_init
Published Specifications
±25 ppm
C/C++ Function Call
Call
niScope_init
following parameters:
resourceName: The device name
assigned by MAX
idQuery:
VI_FALSE
resetDevice:
VI_TRUE
with the
with the