Table 6. NI 5114 Vertical Gain Stimuli, Calibration Test Limits,
Range
Positive Input
(V)
0.04
0.1
0.2
0.4
1
2
4
10
20
40

Programmable Vertical Offset Accuracy

Complete the following steps to verify the programmable vertical offset accuracy for each
device channel.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
18 | ni.com | NI 5114 Calibration Procedure
and Published Specifications
Negative Input
(V)
0.018
0.045
0.09
0.18
0.45
0.9
1.8
4.5
9
18
LabVIEW VI
Calibration
(V)
Test Limits
-0.018
±1.3975%
-0.045
±1.3975%
-0.09
±1.3975%
-0.18
±1.3975%
-0.45
±1.3975%
-0.9
±1.3975%
-1.8
±1.3975%
-4.5
±1.3975%
-9
±1.3975%
-18
±1.3975%
Call
niScope_init
following parameters:
resourceName: The device name
assigned by MAX
idQuery:
resetDevice:
Published
Specifications
±1.5%
±1.5%
±1.5%
±1.5%
±1.5%
±1.5%
±1.5%
±1.5%
±1.5%
±1.5%
C/C++ Function Call
with the
VI_FALSE
VI_TRUE