Trigger Sensitivity

Complete the following steps to verify the trigger sensitivity of the NI 5114. You must verify
channel 0, channel 1, and the external trigger channel using the corresponding iterations listed
in Table 10. Use the following inputs:
For channel 0, use the entries for iterations 1 and 2.
For channel 1, use the entries for iterations 3 and 4.
For the external trigger channel, use the entries for iterations 5 and 6.
32 | ni.com | NI 5114 Calibration Procedure
Range
(MHz)
(V)
125
0.04
125
0.4
125
2
125
10
20
1
C/C++ Function Call
Call
niScope_close
following parameter:
vi: The instrument handle from
niScope_init
Input
Frequency (MHz)
50.1
100.1
50.1
125.1
50.1
125.1
50.1
125.1
20.1
25.1
with the
Published
Specifications
Max Level
Min
(dB)
Level (dB)
1
-1
3
-3
1
-1
3
-3
1
-1
3
-3
1
-1
3
-3
*
N/A
-3
-3
*
N/A