- Product temperature
- Product density
- Product viscosity
- Superimposed pressure
- Sensor installation
- Density selection switch
Measuring accuracy
Deviation
Influence of the process temperature on the switching point
4
3
Fig. 24: Influence of the process temperature on the switching point
1
Shifting of the switching point in mm (in)
2
Process temperature in °C (°F)
3
Switching point at reference conditions (notch)
4
Tuning fork
Influence of the product density on the switching point
SITRANS LVL200S l RELAY (DPDT)
1
10
(
")
25
/
64
8
(
")
5
/
16
6
(
")
15
/
64
4
(
")
5
/
32
2
(
")
5
/
64
0
-2
(
")
-5
/
64
-4
(
")
-5
/
32
-6
(
")
-15
/
64
-8
(
")
-5
/
16
-10
(
")
-25
/
64
0 °C
(32 °F)
+18 ... +30 °C (+64 ... +86 °F)
1 g/cm³ (0.036 lbs/in³) (water)
1 mPa s
0 kPa
vertically from top
> 0.7 g/cm³
± 1 mm (0.04 in)
50 °C
100 °C
150 °C
(122 °F)
(212 °F)
(302 °F)
Supplement
2
200 °C
250 °C
(392 °F)
(482 °F)
27