Instruction Manual October 2004 sitrans LR 400...
Page 2
The user is responsible for all changes and repairs made to the device by the user or the user’s agent. • All new components are to be provided by Siemens Milltronics Process Instruments Inc. • Restrict repair to faulty components only.
Local Programming ..........................22 Auto-Setup ..............................22 Operation ..........................24 General Information ..........................24 Measuring Principle ..........................24 Operating the SITRANS LR 400 ......................26 Selecting a Parameter .........................26 Structure of Parameters ......................27 Changing a Parameter Value ......................27 Disabling and Enabling Programming ....................29 Operating Examples ..........................30 Parameters (HART) ......................32...
Page 4
HART Device Descriptor (DD) ....................92 SIMATIC Process Device Manager (PDM): ................92 HART Communicator 275: ......................93 Appendix VI ........................98 PROFIBUS PA Communications for the SITRANS LR 400 ............98 Device Descriptor ........................98 The GSD file ..........................98 Bus address (Device Address) ....................98 Power Demands ..........................99 Cyclic versus Acyclic Data ......................99...
Please ensure that any safety-related information is confirmed with a qualified Siemens Milltronics representative. WARNINGS: • Changes or modifications not expressly approved by Siemens Milltronics could void the user’s authority to operate the equipment. • This equipment is intended to be used only in fully enclosed metal and concrete containers.
• Installation shall only be performed by qualified personnel and in accordance with local governing regulations. • The SITRANS LR 400 is to be used only in the manner outlined in this manual, otherwise protection provided by equipment may be impaired.
SITRANS LR 400 The SITRANS LR 400 is a long-range FMCW radar level instrument. A liquids version (L) is available for use in liquids storage, performing well on low dielectric liquids or sticky materials requiring a purging unit. A solids version (S) incorporates the Easy Aimer design for use in solids, especially conditions of extreme dust.
Page 8
The end of the antenna must reach inside the vessel through the vessel nozzle (see page 19). If the device is rotated, return the orientation of the housing to its previous position with reference to the flange, to ensure similar performance. Page 4 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
System Implementation SITRANS LR 400 supports HART communication protocol, and SIMATIC PDM software. Typical PLC/mA configuration with HART PLC with mA input card HART HART Communicator 275 PC/laptop with HART modem running PDM S I TR A N S L R 4 0 0...
Specifications Note: Siemens Milltronics makes every attempt to ensure the accuracy of these specifications, but reserves the right to change them at any time. Please ensure these are the most recent specifications. Contact your representative, or check our web site at www. siemens.com/milltronics for the most up-to-date information.
Page 11
Multi–display: 2 freely selectable measured values are displayed simultaneously Programmer (infrared keypad) Siemens Milltronics Infrared IS (Intrinsically Safe) Hand Programmer for hazardous and all other locations (battery is non-replaceable) • approval: ATEX II 1 G, EEx ia IIC T4, certificate SIRA 01ATEX2147 CSA and FM Class I, Div.
Page 12
19.2 kg (42.1 lbs) Easy Aimer with purge option 11.8 kg (26 lbs) Note: Please ensure these are the most recent specifications. Contact your Siemens Milltronics representative, or check our web site at www.siemens.com/milltronics for the most up-to-date information. WARNING: This product is designated as a Pressure Accessory per Directive 97/23/EC and is not intended for use as a safety device.
Page 13
IBM-compatible RAM ≥ 64 Mbytes Hard disk ≥ 100 Mbytes RS 232-C interface VGA graphic card (≥ 640 x 480) • Software for PC/Laptop Windows 95/98/2000/XP or NT 4.0 SIMATIC® PDM 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 9...
Page 14
This product is designated as a Pressure Accessory per directive 97/23/EC and is not intended for use as a safety device. Note: Please ensure these are the most recent specifications. Contact your Siemens Milltronics representative, or check our web site at www.siemens.com/milltronics for the most up-to-date information.
Dimensions SITRANS LR 400 Version for Liquids (without Temperature Extension) 204 mm (8.0") earth connection terminal cover electronics cover 281 mm SITRANS LR 4 00 intermediate (11.1")** flange threaded ring process flange* 191 mm (7.5") 238 mm horn (9.4") 74 mm (2.9")
Page 16
SITRANS LR 400 Version for Liquids with optional Temperature Extension 204 mm (8.0") connection cover electronics cover SITRANS LR 400 intermediate flange 384 mm threaded (15.1") ring process flange* 191 mm (7.5") 238 mm horn (9.4") 74 mm (2.9") 93 mm (3.7")
Page 17
SITRANS LR 400 Version for Solids 204 mm (8.0") connection cover electronics cover SITRA NS LR 400 intermediate 285 mm (11.2") flange 412 mm (16.2") with threaded temperature extension ring Easy Aimer Flange* 9.6 mm (0.38") 188.2 mm horn (7.4") 235.2 mm...
Page 18
The user is responsible for the selection of bolting and gasket materials which will fall within the limits of the flange and its intended use and which are suitable for the service conditions. Page 14 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 19
The user is responsible for the selection of bolting and gasket materials which will fall within the limits of the flange and its intended use and which are suitable for the service conditions. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 15...
Page 20
The user is responsible for the selection of bolting and gasket materials which will fall within the limits of the flange and its intended use and which are suitable for the service conditions. Page 16 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
The customer will supply the purging medium by manual or automatic valve system. This option is only available with universal flanges on the Liquids version of the SITRANS LR 400. 7ML19985FH04 SITRANS LR 400 –...
Installation Notes: • The SITRANS LR 400 is rated for Type 4X/NEMA 4X, Type 6/NEMA 6, IP67. Follow all installation and operating instructions to meet the requirements of this type of protection. Use only approved, suitable sized hubs for watertight applications.
0.07")], as required customer mounting plate with at long horn 9.3"(235 mm) least ø 4" (102 mm) central short horn 7.4"(188 mm) opening, as required 30° Graphic max. labels to be updated 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 19...
4-20 mA, PROFIBUS PA, DC input circuits, 14 - 20 AWG, shielded copper wire • AC input circuit, min 14 AWG copper wire Recommended torque on terminal clamping screws, 0.5 - 0.6 Nm. Page 20 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 25
2G EEx dem [ib] IIC T6 II 2G EEx dem [ia] IIC T6 and II 2G EEx dem [ib] IIC T6 (Liquids version), only the cover of the connection box may be unscrewed for test purposes. The cover on the power supply terminals may not be removed! 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 21...
Then program the unit beginning with the Auto-Setup parameters. Auto-Setup After switching on the SITRANS LR 400, and after a successful self test, press LEFT to access the parameters. Set the Auto-Setup parameters to make the system operational: (see page 33) •...
Measuring Principle The SITRANS LR 400 operates according to the FMCW (Frequency Modulated Continuous Wave) method. Its antenna sends microwaves to the surface of the measuring medium. The wave frequency is modulated continuously (see Determining the Differential Frequency on page 25).
Page 28
Every reflection at a surface generates a different frequency. The reception signal therefore consists of a frequency mix from which the disturbance frequencies must be filtered. These can be caused by fixed targets like struts inside the vessel. Page 25 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
SITRANS LR 400 buttons Selecting a Parameter After a successful self-test, the SITRANS LR 400 displays the two-line multi-display. Press LEFT ARROW to access the parameter menus. The first line of the display tells you the current parameter menu level. The second line shows one of the parameters you can access in the current parameter group.
It does not save a changed setting! For an example of assigning a value from a selection list, see Operating Examples on page 30. Page 27 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 31
If you press UP when the value is at the top of the representable range, SITRANS LR 400 automatically places the value at the next highest position. If 0.9 is displayed and you press UP , the value becomes 1.0. So, 9 becomes 10, 90 or 99 become 100 (depending on whether you have set the input position to the second or first 9), etc.
If Customer Code (Parameter 5.2) is 0, programming of parameters is always enabled. We strongly recommend that a customer code be entered after all programming is completed to secure the programmed values from change. Page 29 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Follow the bold path in the diagram above for input. The other paths lead to other device functions and parameters which are not required in this example. Use the arrow buttons shown next to the numbered operating steps. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 30...
Page 34
Select the last decimal place with RIGHT arrow And end the input with RIGHT arrow (ENTER function). Page 31 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
They apply to all applications and are required to make the system operational. 1. Auto-Setup 1.1: Language Local (F = English) Language of the local user interface English Value Deutsch Page 33 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 37
Set the LRV as the level above the bottom of the vessel (see Functional Dimensions on page 33) in the units system selected with Function 4.1. 1 . It corresponds to an output current of 4 mA. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 34...
Display of two measured values. Values are determined in Parameter 4.5. 1 .1 (Line 1 Local) and in Parameter 4.5.1.4 (Line 2 Local). 2.2: Level (F = m) Current level of measured medium (set unit using Length Unit in Auto-Setup) Page 35 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 39
3.1.1.3: Minimum Temperature (F = 26 °C) Previous minimum internal temperature of the device 3.1.1.4: Aging Approximate value for the previous life of the device in % (100% = approx. 10 years) 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 36...
Page 40
3.1.x: Software Diagnostic messages of the software and/or 3.1.x: Application Diagnostic messages to the application and/or 3.1.x: Parameters Display of the false parameters and/or 3.1.x: Service For service purposes only Page 37 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 41
When this parameter is accessed and a value is entered, the device sets the defined current value that can be validated. Complete the parameter function by pressing LEFT so the analog output again gives the measured value. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 38...
Page 42
S/N ratio provides a relative measure of the strength of reflection of the measuring medium in dB. Its display can be evaluated as follows: • x > 20: very good • 20 > x > 10: good • x < 10: satisfactory Page 39 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 43
4.1.1: Length Unit = Parameter 1.2 4.1.2: Volume unit (F = m bush Value bbl (liq) ImpGal 4.1.3: Mass Unit (for qualified personnel only) This parameter is for Factory Authorized Personnel Only. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 40...
Page 44
By selecting yes or no, you specify whether the device is mounted on a stilling pipe. If you select yes, Parameter 4.2.1.3.2 is enabled so you can specify the internal diameter of the stilling pipe. Page 41 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 45
If your measuring medium forms waves more than 1 cm in height, you should select the wavy setting. The turbulent setting is recommended for waves greater than 10 cm. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 42...
Page 46
This assigns a greater probability to measuring targets which move at this speed. Page 43 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 47
4.2.3: Sensor Parameter Here you can view and change the sensor parameters which you have selected according to Parameter 4.2.2 (Measuring Conditions). Note: The factory settings for the user vessels are not editable. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 44...
Page 48
Dynamic processes in the vessel are included in the evaluation of the measuring targets. The typical filling speed can be set in Parameter 4.2.2.5. If the measured value still does not follow the level height, switching off echo motion may improve the result. Page 45 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 49
Switch off window tracking for applications where the SITRANS LR 400 is unable to keep up to level changes. By using SIMATIC PDM, you can display a list of all echoes in your vessel. It provides the distance between the flange and the measuring medium’s surface, as well as the distances of fixed targets.
Page 50
Defines (in percent) how high the TVT (Time Varying Threshold) curve is placed above the profile, relative to the largest echo. When SITRANS LR 400 is located in the center of the vessel, lower this parameter to prevent multiple echo detections.
Page 51
Set the hysteresis of the limit values in the units system selected according to Parameter 4. 1 .1 (see diagram below). Limit value alarm measuring variable measuring variable upper limit value hysteresis hysteresis limit value lower limit alarm value limit value alarm time time 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 48...
Page 52
4.2.5.6: HYST Volume (F = 0.5 m Hysteresis of the volume limit values Value numerical value 4.2.5.7: Tank Characteristic (F = Calculate) Determining the vessel characteristic Calibrate/table Value Calculate Page 49 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 53
The 4.2.5.8 Calibrate/table parameter offers the following selection possibilities: 4.2.5.8.1: Calibrate Here you can enter up to 50 reference values whose levels SITRANS LR 400 measures. Enter the appropriate volume (determined by manual calibration). If you access this parameter, first the currently measured level is displayed. Accept it by pressing LEFT .
Page 54
≤ 1 % may occur. As well, the necessary data which you can get from the design documents of your vessel must still correspond to the real conditions. The 4.2.5.8: Calculate parameter requires the following parameters: Page 51 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 55
4.2.6.3: Mass Damping (for qualified personnel only) 4.2.6.4: MinLim Mass (for qualified personnel only) 4.2.6.5: MaxLim Mass (for qualified personnel only) 4.2.6.6: HYST Mass (for qualified personnel only) 4.2.6.7 Tank Characteristic (for qualified personnel only) 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 52...
Page 56
4.2.8.x: User2 Parameters (for qualified personnel only) All parameters in range 4.2.8.x are for Factory Authorized Personnel Only. Advanced echo processing parameters should only be modified by qualified personnel. Page 53 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 57
Upper current limit Value 20 - 22.5 mA Here you can set the upper current limit of the output signal in steps of 0. 1 mA (see Current limiting diagram on next page). 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 54...
Page 58
If the fault persists after the set time runs out, the analog output switches to an error signal of 3.6 mA. Note: A fault is different from a loss of echo which indicates a failsafe condition. Page 55 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 59
When D is selected, all errors are displayed. When D+F is selected, there is special handling for failsafe. When D+F+W is selected, there is special handling for warnings. 4.4.3: Signal Type DO (F = Relay closes) Relay closes Value Relay opens 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 56...
Page 60
Choice of measured value in line 1 Level Value Volume Mass Note: The value of 4.5.1.3 is determined by the selection (Level, Volume or Mass) in Parameter 4.5.1.1 (Line 1 Local). Page 57 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 61
Parameter 4.5.1.5 is displayed if Parameter 4.5. 1 .4 (Line 2 Local) = Level, Volume, or Mass. If 4.5.1.4 (Line 2 Local) = Level 4.5.1.5: Level Parameter (= Parameter 4.2.4) If 4.5.1.4 (Line 2 Local) = Volume 4.5.1.5: Volume Parameter (= Parameter 4.2.5) 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 58...
Page 62
4.6.4: Antenna and Flange 4.6.4.1: Flange Size (according to customer specifications) Size of the flange DN 50, 2" DN 80, 3" Value DN 100, 4" DN 150, 6" Special design Page 59 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 63
4.6.4.6: Flange Material (according to customer specifications) Value 316/316L Special design 4.6.4.7: Seal Material (according to customer specifications) Sealing material PTFE Value 4.6.5: Tag (according to customer specifications) Device identification Value up to any eight characters 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 60...
Page 64
Distance from sensor to flange The antenna offset defines the propagation time of the measuring signal between the sensor and the flange as a distance. It is preset at the factory and cannot be changed. Page 61 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 65
5.3: Factory Reset (F = no) Reset all parameters to factory setting Value This parameter allows you to reset all parameters to the original factory setting as described in Parameters (HART) beginning on page 32. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 62...
Display of two measured Level/Signal- non-editable values to-noise ratio in dB 2.2: Level Level of measured non-editable medium 2.3: Volume Volume of measured non-editable medium 2.4: Mass For qualified personnel only Page 63 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 67
3.1.x: Parameters parameters and/or 3.1.x: Service For qualified personnel non-editable only 3.2: Device test 3.2.1: Self-test Check device state non-editable 3.2.2: Display test Visual check of LCD non-editable 3.3: Sensor variables 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 64...
Page 68
Units system for all other SI unit units US/UK unit 4.2: Operating parameters 4.2.1: Tank geometry 4.2.1.1: Nozzle height = [1.3] 4.2.1.2: Tank height = [1.4] 4.2.1.3: Stilling pipe? Stilling pipe available? [Liquids} Page 65 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 69
4.2.2.8: Failsafe timer Sets the time delay, in 10 min 1 min minutes, before going 2 min into fail-safe level etc. 4.2.2.9: Range exten- Sets the distance below sion the tank height included in the evaluation etc. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 66...
Page 70
4.2.3.B Window Defines position of win- 0 to 100% Trigger dow on leading edge of selected echo, relative to amplitude 4.2.4: Level param. 4.2.4.1: Level URV = [1.5] Page 67 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 71
4.2.5.8: HYST volume Hysteresis of the volume numerical value 0.5 m limit values 4.2.5.9: Tank charac- Determining the tank Calculate Calibrate/table teristic characteristic Calculate 4.2.5.A: Calibrate/table 4.2.5.A: Calculate 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 68...
Page 72
4.2.6.8: HYST Mass For qualified personnel only 4.2.6.9: Tank Charac- For qualified personnel only 4.2.6.A: Calibrate/table 4.2.6.A: Calculate 4.2.6.A.1: Calibrate For qualified personnel only For qualified personnel only 4.2.6.A.1: Tank design Page 69 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 73
1 Volume Mass 4.4.1.2: Display local Method of display in line Eng unit Eng unit Bargraph 4.4.1.3: Level param. = [4.2.4] 4.4.1.3: Volume = [4.2.5] param. 4.4.1.3: Mass param. = [4.2.6] 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 70...
Page 74
4.5.4: Antenna and flange 4.5.4.1: Flange size Rated size of the flange according to DN 50, 2" customer DN 80, 3" specifica- DN 100, 4" tions DN 150, 6" Special design Page 71 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 75
32 char- sage, e.g. the date of last customer acters check or clean specifica- tions 4.5.8: Manufacturer identification 4.5.8.1: Serial no. Factory serial number unique num- non-editable 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 72...
Page 76
Input of customer code to Customer code enable programmability 5.2: Customer code Determination of cus- up to 9 digit code tomer code 5.3: Factory reset Reset all parameters to factory setting Page 73 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Please review the instructions below for troubleshooting before contacting the responsible customer services. Classification of Faults Faults occurring in the SITRANS LR 400 can be classified in the following groups: • faults caused by ambient influences: over- and under-temperature, moisture,...
Liquid Store application type 4.2.2.1. instead of first echo. SITRANS LR 400 reads low or empty when Aim the SITRANS LR 400 using shims or order an material level is high Easy Aimer kit. Contact your Siemens Highly sloped surface representative for service and possible User parameter support.
Please note the following: • The user is responsible for all changes and repairs made to the device. • All new components must be provided by Siemens Milltronics Process Instruments Inc. • Restrict repair to faulty components only •...
If the antenna is contaminated, any built-up material can be removed using air or soft brushes. Be careful not to damage the PTFE emitter cone inside the antenna. An optional purge kit is available from Siemens Milltronics if required. Contact your local Siemens Millltronics representative for more details.
Metal instrument formed as a cubic segment with right angles. Upper limit of the valid measuring range as a distance from the bottom inside of the vessel. Validity Measure of the certainty of the current measured value in %. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 79...
4.2.6.8.4 Correction factor 4.2.2.4 Current limit 4.3.4 Current output Customer code Dead band 4.2.2.3 Descriptor 4.6.6 Device data Device Information Device revision 4.6.8.3 Device test Diagnostics Digital output Digital output Display Page 80 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 83
Level parameter (line 1 local) 4.5. 1 .3 Level parameter (line 2 local) 4.5. 1 .5 Level URV Level URV 4.2.4.1 Line 1 local 4.5. 1 .1 Line 2 local 4.5. 1 .4 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 81...
Page 84
Range Extension 4.2.2.9 Raw value 3.4. 1 Reference distance 4.6.8.7 Reflectivity 4.2.2.6 S/N ratio 3.4.3 Seal material 4.6.4.7 Self-test 3.2. 1 Sensor damping 4.2.3.1 Sensor parameter 4.2.3 Sensor temperature 3.4.5 Sensor variables Page 82 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Length unit 4. 1 .2 Volume unit 4. 1 .3 Mass unit 4. 1 .4 Temperature unit 4. 1 .5 Other units 4.2.1.1 Nozzle height 4.2.1.2 Tank height 4.2.1.3 Stilling pipe? Page 84 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Line 2 local 4.5.1.5 Level parameter 4.5.1.5 Volume parameter 4.5.1.5 Mass parameter 4.5.2 Language local 4.5.3 LCD backlight 4.6.1 Power supply 4.6.2 Process temperature 4.6.3 Electrical connection 4.6.4 Antenna and flange Page 86 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Warning: Internal temperature must not exceed 85 °C (185 °F)! process temperature Warranty may be void. Page 88 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
* standard process seal is rated to a max. of 200 °C of continuous duty. **process seal for hazardous location is rated to a max. of 250 °C of continuous duty. 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 90...
Page 92
• Never attempt to loosen, remove, or disassemble process connection or instrument housing with vessel contents. • Improper installation may result in loss of process pressure. Page 91 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
HART and PROFIBUS PA devices. The HART DD for the SITRANS LR 400 was written with SIMATIC PDM in mind and has been extensively tested with this software. The HART DD for the SITRANS LR 400 is available on the SITRANS LR 400 page of our website at www.siemens.com/milltronics.
1. Serial number 5. Fld. dev. rev. 2. Device order number 6. Software rev. 3. Actual order number 7. Hardware rev. 4. Software version 5. Hardware version 6. Antenna offset 7. Reference distance Page 93 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
3. Mass damping 4. Clear table 4. MinLim Mass 5. MaxLim Mass calculate 6. HYST Mass 1. Tank design 7. Tank characteristic 2. Bottom design 8. Calibrate/table 3. Tank volume or Calculate 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 94...
1. Eng unit 2. % 3. Bar graphic Line 2 Local 1. Level 2. Volume 3. Mass 4. Temperature 5. Validity 6. S/N ratio 7. Amplitude 8. Digital output 9. Analog output Page 95 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Page 97
Supported HART Commands: The SITRANS LR 400 conforms to HART rev. 5 and supports the following: Universal Commands 0, 1, 2, 3, 6, 11, 12, 13, 14, 15, 16, 17, 18, 19 Common Practice Commands 33, 34, 35, 36, 37, 38, 40, 41, 42, 43, 44, 48, 50, 51, 53, 59, 108, 109, 110...
Page 98
The HART commands are rarely if ever used by end users. For details on the Universal and Common Practice Commands, please contact the HART Communication Foundation. For details on the Device Specific Commands, please contact Siemens Milltronics. Page 97 SITRANS LR 400 – INSTRUCTION MANUAL...
PROFIBUS PA is an open industrial protocol. Full details about PROFIBUS PA can be obtained from PROFIBUS International at www.profibus.com The SITRANS LR 400 is a Class A, Profile Version 3.0, PA device. It supports Class 1 Master for Cyclic data exchange, and Class 2 for acyclic services: (See below for details).
Input and output information is always requested at every bus scan and is set up as cyclic data. Configuration information is only needed periodically and is set up as acyclic data. Cyclic Data When you configure the SITRANS LR 400 on the PROFIBUS PA bus, there are two slots available for modules. Note: Each of the slots has to have a module defined in it.
Page 101
Floating Point Status byte1 byte 2 byte 3 byte 4 byte 5 Level Free — — — — — Place Mass byte 6 byte 7 byte 8 byte 9 byte 10 7ML19985FH04 SITRANS LR 400 – INSTRUCTION MANUAL Page 100...
0x40 non-specific 0x8E high limited active critical alarm 0x8D low limited active critical alarm 0x8A high limited active advisory alarm 0x89 low limited active advisory alarm 0x84 active update event 0x80 Page 101 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...
Import the GSD file SM_062A.GSD from the Siemens Milltronics Web site: www.siemens.com/milltronics into Step 7 software. Add the SITRANS LR 400 “rack”: click and drag the SITRANS LR 400 folder from the hardware catalog. Fill the rack with desired modules, by dragging and dropping them from the hardware catalog.
Hazardous Installation The necessary certificates are enclosed separately. Please refer to the SITRANS LR 400 Certificate manual 7ML19985FP82. Page 103 SITRANS LR 400 – INSTRUCTION MANUAL 7ML19985FH04...